Experimental determination of beam quality correction factors in scanned proton beams using water calorimetry
Lies Verpoest, Séverine Rossomme, Kevin Souris, Peter Kuess, Hugo Palmans, Stefano Lorentini, John A. Lee
European Society for Therapeutic Radiology and Oncology · Ion Beam Applications (Belgium)
Key points
From the paper's conclusion- This study presents new experimental <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" overflow="scroll"> <mml:mrow> <mml:mrow> <mml:msub> <mml:mi>k</mml:mi> <mml:mi>Q</mml:mi> </mml:msub> </mml:mrow> </mml:mrow> </mml:math> -factors for ionization chambers in single-layer scanned proton beams, contributing to the currently scarce experimental database, providing further validation of the T
Abstract
Abstract Objective. To measure beam quality correction factors ( k Q ) in single-layer scanned proton beams using water calorimetry for three ionization chamber types commonly used in clinical proton dosimetry. Approach. Measurements were performed at two proton therapy centers using clinical proton beams with nominal energies of 150, 220 and 226 MeV, at a reference depth of 4 cm. The k Q -factors were obtained by comparing absorbed dose-to-water determinations from a water calorimeter with ionization chamber readings under identical conditions. Three ionization chamber models were investigated: the IBA FC65-G (cylindrical), PPC05 and PPC40 (plane-parallel). Two independent calorimeter setups were used across three measurement campaigns. Main results. The measured k Q -factors showed strong agreement with current TRS-398 Rev. 1 (2024) recommendations and literature data. For the FC65-G chamber, excellent alignment was observed with previous water calorimetry-based measurements. For the PPC40 chamber, both chambers yielded consistent results within 0.5% of TRS-398 Rev. 1. For the PPC05 chambers, a maximum deviation of 1.4% was observed relative to TRS-398 Rev. 1, and inter-chamber variability was within 0.5%. The use of two calorimeter setups yielded consistent results within 0.2%, supporting their validity. Significance. This study presents new experimental k Q -factors for ionization chambers in single-layer scanned proton beams, contributing to the currently scarce experimental database, providing further validation of the TRS-398 Rev. 1 (2024) recommendations, and supplying benchmark data for future Monte Carlo-based k Q calculations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimentally derived beam quality correction factors for PPC05 and razor nano ionization chambers in proton scanned beams based on graphite calorimetry.Journal of applied clinical medical physics · 2026
- Monte Carlo calculated beam quality correction factors for light ions used in particle therapyPhysics in Medicine and Biology · 2026
- Ion recombination correction in reference dosimetry for pencil beam scanned proton beamsMedical Physics · 2026
- Three-dimensional computational evaluation of water-to-air mass stopping power ratio for therapeutic helium-, carbon-, oxygen-, and neon-ion beams and its application to patient-specific quality assurance.Physics in medicine and biology · 2026
- Impact of a common user-derived electrometer calibration coefficient on absorbed dose-to-water determination: a multicenter uncertainty studyJournal of Radioanalytical and Nuclear Chemistry · 2026
- Development and evaluation of a water-profile-based calibration method for 2D ionization chamber arrays in medical linear accelerator quality assurance.Journal of applied clinical medical physics · 2026