Evaluation of registration accuracy using short-arc cone-beam computed tomography for image-guided radiotherapy in breast cancer.
Shuhei Wada, Masahide Tominaga, Motoharu Sasaki, Koichi Yamanashi, Yoshio Iwai, Shunsuke Furutani, Yusei Shinohara, Yoshihiro Uto
Graduate School of Advanced Science and Technology, Tokushima University, Tokushima, Tokushima, Japan. · Graduate School of Biomedical Sciences, Tokushima University, Tokushima, Tokushima, Japan.
Key points
From the paper's conclusion- These findings suggest its clinical applicability with appropriate acquisition angle settings.
Abstract
BACKGROUND: Image-guided radiotherapy (IGRT) using cone-beam computed tomography (CBCT) enables highly accurate imaging registration; however, increased radiation exposure remains a concern. Recently, short-arc CBCT (SA-CBCT), which reduces projection data and reconstructs images, has become available. Although previous studies have shown that image quality deteriorates when projection data are reduced, no study has directly evaluated the clinical usefulness of SA-CBCT or its effect on registration accuracy. PURPOSE: We verified the registration accuracy of SA-CBCT in breast tangential irradiation using an anthropomorphic phantom and clinical data, and examined its clinical feasibility compared with the commonly used 200° scan. METHODS: An anthropomorphic phantom was placed on the treatment bed, and CT and CBCT images acquired during treatment planning were aligned. Bed movement was measured for both conventional CBCT and SA-CBCT images. The difference between the two measurements was quantified as N, defined as the norm of the vector difference between CBCT and SA-CBCT registration shifts, to evaluate registration accuracy. The acquisition angle was varied to assess the effects of central and total angles. In the clinical study, registration data from 30 consecutive patients who underwent whole-breast irradiation were analyzed. Data were reconstructed by changing the acquisition angle from 200° to 120°, and N was calculated as in the phantom experiment. RESULTS: In the phantom experiment, N was within 0.15 mm (3-axis) and 0.55 mm (6-axis) when the acquisition angle was reduced to 120°. In clinical data, the average N was 0.64 mm in 3-axis registration (84.7% within 1 mm) and 0.85 mm in 6-axis registration (71.3% within 1 mm). Even with a 120° acquisition angle, the body contour and chest wall edges were sufficiently preserved. CONCLUSIONS: SA-CBCT maintained adequate registration accuracy in breast tangential irradiation. These findings suggest its clinical applicability with appropriate acquisition angle settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy.Diagnostics (Basel, Switzerland) · 2026
- Manual CBCT registration by Japanese radiological technologists: agreement with a clinical reference position and exploratory associations with professional experience.Radiological physics and technology · 2026
- Beyond CBCT: real-time motion management in spine and non-spine bone SBRT using stereoscopic X-rays and surface guidance.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Elliptical acquisition provides measurement accuracy comparable to CT in an experimental model of hip surgery.Orthopaedics & traumatology, surgery & research : OTSR · 2026
- Development of a Statistical Framework for Optimizing Image-Guidance Frequency in Whole-Breast Radiotherapy.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Field of View Optimization in Cone-Beam Computed Tomography for Bilateral Alveolar Clefts Before and After Bone Grafting.The Cleft palate-craniofacial journal : official publication of the American Cleft Palate-Craniofacial Association · 2026