TotalSegmentator Aorta Report: Automated Aortic Diameter Measurement at 11 Standardized Landmarks on CT Images
Jakob Wasserthal, Daniel Hinck, Martin Segeroth, Joshy Cyriac, Denis Berdajs, Jules Miazza, Maurice Pradella
Key points
From the paper's conclusion- Conclusion: The tool measured aortic diameters at 11 landmarks along the entire aorta with an MAE of about 1 mm, below published interreader variability.
Abstract
Purpose: To develop and evaluate a fully automated, open-source tool that measures aortic diameters at 11 standardized landmarks from the aortic annulus to the bifurcation on any CT image, following current guideline recommendations. Materials and Methods: This retrospective single-center study included 239 arterial phase CT angiography examinations of 230 patients (median age, 65 years [range, 18-99 years]; 163 men) performed between April 2025 and October 2026 (209 electrocardiographically [ECG] gated). The tool locates landmarks from TotalSegmentator segmentations of aortic branches and vertebrae and measures the maximum diameter at each. A senior radiologist reviewed and, where needed, corrected the automated measurements to create the reference standard. Agreement was assessed with mean absolute error (MAE), bias, and differences greater than 2 mm. Results: Across all landmarks, MAE was 1.06 mm (95% CI: 0.91, 1.22), and 86.6% (95% CI: 84.4, 88.5) of differences were 2 mm or less. The tool measured slightly larger diameters (bias, 0.45 mm [95% CI: 0.34, 0.57]). The sinuses of Valsalva had the largest MAE (2.17 mm [95% CI: 1.59, 2.86]). For 1.2% of landmarks, the automatic measurement failed. Examination-level MAE did not differ significantly by pathology, atherosclerosis grade (P=.95), or ECG gating (P=.14). Examinations with mural thrombus more often had an MAE greater than 2 mm (50% [95% CI: 31, 69] vs 20%; P=.03). Conclusion: The tool measured aortic diameters at 11 landmarks along the entire aorta with an MAE of about 1 mm, below published interreader variability. Agreement was lower at the aortic root and with mural thrombus. Code availability: The tool is open source and freely available at https://github.com/wasserth/TotalSegmentator.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-centre Validation of Automated Outer-to-Outer Wall Segmentation of Infrarenal Abdominal Aortic Aneurysms Using a nnU-Net Based Model.Journal of imaging informatics in medicine · 2026
- AI-Driven CT Analysis for Aortic Disease: A Task-Based Systematic Review of Segmentation, Measurement, and Detection Applications.Tomography (Ann Arbor, Mich.) · 2026
- CT-derived LVOT diameter refines aortic stenosis assessment in TAVI candidates with suboptimal echocardiographic image qualityEuropean Heart Journal · 2026
- Automated CTA-Based Assessment of Aortic Remodeling Through Volumetric Biomarkers in Type B Aortic DissectionmedRxiv · 2026
- AI-based Framework for Automated Segmentation and Surgical Decision Support in Ascending Thoracic Aortic Aneurysm.IEEE journal of biomedical and health informatics · 2026
- Artificial Intelligence in 3D Aortic CTA Segmentation for Abdominal Aortic Aneurysm Management: A Narrative Review.Journal of imaging · 2026