Design and Performance Evaluation of a Quad-panel Dual-isotope PET Imaging System
Yu Shi, Tengjiao Wang, Sen Yang, Bo Wen, Meng Niu, Yirong Wang, Risheng Yang, Yanyun Liu,
Key points
From the paper's conclusion- We designed and evaluated a compact quad-panel dual-isotope PET system without dedicated prompt gamma detectors.
Abstract
In simultaneous dual-isotope PET, prompt-gamma-based separation is challenged by the low statistics of triple coincidences and the limited flexibility of coincidence logic for multi-energy event extraction. We designed and evaluated a compact quad-panel dual-isotope PET system without dedicated prompt gamma detectors. The system comprised four orthogonally arranged detector panels, each providing an effective area of 107.0 * 145.5 mm. GPU-accelerated software coincidence processing extracted double- and prompt-gamma-tagged triple coincidences from raw single-photon data. Triple-coincidence images were reconstructed with correlated-random-triple correction using the consistent delayed expectation-maximization method, followed by image-domain isotope separation using a spatially varying scaling factor. System performance was evaluated using phantom and in vivo studies, with emphasis on 68Ga/18F imaging because of the low prompt gamma yield. The system achieved energy resolutions of 15.7%, 12.9% and 10.2% at 511, 1077, and 1275 keV, and a timing resolution of 1.35 ns. In the 68Ga Derenzo, 1.6-mm rods were resolved in double-coincidence images, with 1.2-mm rods partially distinguishable, while 1.6-mm rods remained discernible in triple-coincidence images. The mean prompt gamma detection efficiencies along the three axes ranged from 8.19% to 9.31%. The system sensitivity was 1.12% for doubles with an energy window of 350-650 keV, and 0.068% for triples after random correction with a dual-energy window. The scatter fraction was 27.81%, and the NECR reached 83 kcps at 9.38 MBq for doubles and 106 cps at 21.42 MBq for triples. Phantom and in vivo studies demonstrated successful separation of 18F/68Ga and 18F/124I tracer distributions. By maintaining high-resolution and quantitative imaging performance, the system provides a preclinical platform for multitracer PET studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Joint Poisson Reconstruction for Prompt-Gamma-Enabled 68Ga/18F Dual-Isotope Quantitative PET ImagingarXiv · 2026
- Benchmarking 18F and 68Ga Phantom Performance on a Digital Time-of-Flight PET/CT System: Reference Values for Spatial Resolution, Contrast Recovery, and Background Variability.Journal of nuclear medicine technology · 2026
- Towards real-time ion range verification via hybrid Compton-PET 3D imaging at isochronous cyclotronsarXiv · 2026
- Regularization in multiplexed PET image reconstruction.Medical physics · 2026
- Influence of Radionuclide Cascade Decay Characteristics on Non-Collinear Coincidence Imaging Performance: A GATE-Based Monte Carlo Simulation Study.Biomedical physics & engineering express · 2026
- A high-resolution TOF-DOI PET system for nonhuman primate neuroimagingEJNMMI Physics · 2026