arXivPreprintOpen accessphysics.ins-det

CMS ECAL VFE Production Qualification

C. Borca, S. Argir\`o, L. T. Bulaja, D. Brunet, F. Cavallari, D. Cristoforetti, G. Dissertori, L. Finco,

Key points

From the paper's conclusion
  • We have received and functionally tested 13596 unique VFEs; two never passed our test criteria.

Abstract

Upgrading the CMS ECAL barrel for the HL-LHC requires replacement of all on-detector electronics, including 12240 Very-Front-End (VFE) cards employing CATIA and LiTE-DTU ASICs for amplification and digitisation of signals from ECAL's 61200 lead-tungstate crystals. All production VFEs undergo the same qualification procedure: functional testing, environmental stress screening, and calibration. We have received and functionally tested 13596 unique VFEs; two never passed our test criteria. Results for other VFEs are uniform and within expectations, demonstrating the excellent quality of the cards. Key metrics, such as a gain ratio of $10.21 \pm 0.07$ and power consumption of $1.805 \pm 0.013~\mathrm{W}$, demonstrate high consistency across the batch.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.