Developing the NewAthena X-IFU Cryogenic AntiCoincidence Detector (CryoAC): From Microfabrication Process Standardization to Cryogenic Functional Verification Toward TRL5.
Claudio Macculi, Matteo D'Andrea, Giacomo Gorla, Simone Lotti, Gabriele Minervini, Francesco Monastra, Luigi Piro, Lorenzo Ferrari Barusso,
INAF/IAPS Roma, Via del Fosso del Cavaliere 100, 00133 Rome, Italy. · INAF HQ, Viale del Parco Mellini 84, 00136 Roma, Italy.
Key points
From the paper's conclusion- This work presents the complete development cycle of the Demonstration Model 1.2 (DM 1.2), which is aimed at completing the TRL5 demonstration path featured by all the critical technologies operating simultaneously.
Abstract
The Cryogenic Anticoincidence (CryoAC) detector is a critical subsystem designed to reduce the particle background for the X-ray Integral Field Unit (X-IFU) instrument onboard the NewAthena space observatory, the next ESA X-ray Large mission. Advancing this technology to Technology Readiness Level 5 (TRL5) requires a unified validation spanning both cleanroom microfabrication repeatability and mK low temperature operational performance. This work presents the complete development cycle of the Demonstration Model 1.2 (DM 1.2), which is aimed at completing the TRL5 demonstration path featured by all the critical technologies operating simultaneously. First, single-process verification protocols were established for Iridium pulsed laser deposition, Reactive Ion Etching (RIE), and deep silicon trenching via the Bosch process. Second, three identical single-pixel prototypes were fabricated and subjected to mK characterization. Four-wire resistance measurement results confirmed a 2/3 production yield against strict design targets (TC ~100 mK). Finally, functional testing at a bath temperature of 50 mK using a VTT FAB4 SQUID readout demonstrated excellent performance, including a low-energy threshold of ~0.6 keV, a pixel power dissipation of 5.15 nW, and an energy resolution ΔEFWHM = 735 eV at 6 keV. These combined achievements successfully validate the entire manufacturing and operational baseline against all primary space mission requirements. This paper has to be considered as a review of the CryoAC technology path toward the TRL5 achievement; main findings will be reported and discussed. Details are relegated to other papers.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Characterization of Silicon-Membrane TES Microcalorimeters for Large-Format X-ray Spectrometers with Integrated Microwave SQUID Readout.IEEE transactions on applied superconductivity : a publication of the IEEE Superconductivity Committee · 2026
- In-flight performance of the XRISM/Resolve detector systemJournal of Astronomical Telescopes Instruments and Systems · 2025
- A cryogenic characterization platform for commercial off-the-shelf (COTS) electronic devices in liquid xenon detectors for high-energy physics experiments.The Review of scientific instruments · 2026
- CHARMS250V1: A Cryogenic Front-End ASIC for Low-Noise Readout of Charge or Light Signals in Noble Liquid DetectorsarXiv · 2026
- The CosmicWatch Desktop Muon Detector (v3X): Prototype Design and Adaptation for SpaceflightarXiv · 2026
- Development of millikelvin scintillation photodetectors with CUPIDarXiv · 2026