Study of three-dimensional dose distribution measurement of ultra-high dose rate electron beam.
Wei Zhang, Chang Cheng, Ronghu Mao, Leijie Ma, Dehong Li, Zhihua Peng, Keying Zhu, Wenjun Liu,
Soochow University, Soochow University, Suzhou, Jiangsu, 215006, China. · Beihang University, Beihang University, XueYuan Road No.37, HaiDian District, BeiJing, China, Beijing, 100091, China.
Key points
From the paper's conclusion- Calculations of the gamma passing rate on each plane of the reconstruction result indicate that the method used shows high accuracy for the transverse distribution. 
Significance: The proposed three-dimensional dose distribution measurement method is applicable to both conventional dose rate radiotherapy and UHDR radiotherapy conditions.
Abstract
FLASH radiotherapy which employs an ultra-high dose rate (>40 Gy/s), effectively protects normal tissues and organs while maintaining robust tumor control. However, it also presents serious challenges for quality assurance (QA). Currently, dosimeters face various limitations under FLASH radiotherapy conditions which make it difficult to meet the demand for rapid and high-precision three-dimensional dose verification. In this paper, CMOS cameras were used to capture luminescence images generated by irradiated water and for three-dimensional dose reconstruction. Approach: A tri-orthogonal view camera imaging setup was designed and constructed to collect radioluminescence signals in water phantom. The ordered subsets expectation maximization (OSEM) iterative algorithm was adopted to obtain three-dimensional dose reconstruction. The experiments were performed by 6 MV and 10 MV conventional dose rate X-rays, as well as 9 MeV electron with an ultra-high dose rate (UHDR) of 120 Gy/s. The dose distribution was verified by treatment planning system calculations and film measurements, respectively. Quantitative evaluations were performed through percentage depth dose (PDD), lateral dose profiles, and two-dimensional gamma analysis (3%/3 mm, 10% threshold). Results: The luminescence intensity of water phantom shows high linearity with the irradiation dose and exhibits independent characteristics with dose rate. The PDD curve of conventional X-ray reconstruction underestimates the dose in the dose fall-off region, and the transverse dose distribution shows high consistency with the reference distribution. Under different irradiation fields, the difference in the full width at half maximum (FWHM) of the lateral dose distribution curves at each depth is less than 1.25 mm. The PDD curve of the UHDR electron beam reconstruction result underestimates the dose in the shallow region. The deviation of the half-peak dose depth from the film measurement result is less than 0.7 mm. Calculations of the gamma passing rate on each plane of the reconstruction result indicate that the method used shows high accuracy for the transverse distribution. Significance: The proposed three-dimensional dose distribution measurement method is applicable to both conventional dose rate radiotherapy and UHDR radiotherapy conditions. It shows high accuracy in the evaluation of lateral parameters such as field shape, field size, flatness and uniformity, and can meet the basic requirements for field consistency verification in clinical UHDR radiotherapy QA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Toward quantitative reconstruction of delivered dose in a 3D digital water phantom for electron FLASH-RT via annular-array electron-induced acoustic imaging.Physics in medicine and biology · 2026
- Pulse-resolved spatial dosimetry of ultra-high dose rate electron FLASH beams using optical imaging.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Radiation monitoring in the scattered field of a clinical electron FLASH beam with Timepix4Journal of Instrumentation · 2026
- Ultra-high-dose-rate pulsed irradiation and the conditional FLASH effect: dosimetry, multiscale biology, and clinical translation.Frontiers in oncology · 2026
- Evaluation of the Exradin A30 Parallel Plate Ion Chamber as a Reference Dosimeter in Ultra-High Dose Rate (UHDR) Electron Beams.ArXiv · 2026
- Dosimetric Characterization and Workflow Optimization of the FLASH-SARRP for Reliable Preclinical Radiobiological StudiesbioRxiv · 2026