Power Consumption of Picture Archiving and Communication System (PACS) Workstations: Quantification and Comparison of Operational Modes.
Anna Luisa Kühn, Domenic Mongillo, Bilal El Kaddouri, Seyedeh Panid Madani, Daria Kifjak, Alexander A Bankier
Department of Radiology, University of Massachusetts Chan Medical School, Worcester, USA. · Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Vienna, AUT.
Key points
From the paper's conclusion- PACS workstations represent a substantial and likely underrecognized source of power consumption in radiology. Quantitative insights and simple decision tools, such as the conversion factors and break-even times presented in this study, can support more energy-efficient workstation management and reduce the environmental impact of radiology operations.
Abstract
PURPOSE: This study aimed to quantify the power consumption of Picture Archiving and Communication System (PACS) workstations across common operating modes. MATERIALS AND METHODS: Power consumption was measured for three PACS workstation setups (two hospital-based, one home-based) across five operating modes, including active, standby, powered-down, and transition states. Measurements were conducted using an external power meter, with three repeated measurements per mode and setting (total N = 45). Static modes were extrapolated from 10-minute intervals; dynamic modes included measured duration. Statistical analysis assessed variability and differences between settings. Departmental consumption was estimated by scaling per-workstation data to 105 units and benchmarked against common household appliances. Conversion ratios and break-even times for transitioning between modes were calculated. RESULTS: Measurement variability was low, with no significant differences between repetitions or inclusion of mammography workstations. At the departmental level, the total consumption reached 24.99 kW in full operation. Even non-active states had considerable power consumption (18.37 kW logged off; 2.41 kW powered down). Mode comparisons showed up to 7.61-fold differences in power consumption. Break-even times were 6.67 minutes (logged-off to active) and 5.86 minutes (powered-down to active), indicating short idle periods justify reduced-power states. CONCLUSION: PACS workstations represent a substantial and likely underrecognized source of power consumption in radiology. Quantitative insights and simple decision tools, such as the conversion factors and break-even times presented in this study, can support more energy-efficient workstation management and reduce the environmental impact of radiology operations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying energy savings in radiology: a simple approach to make the radiology department more sustainableBMJ Leader · 2025
- Energy sustainability of breast ultrasound: comparison of automated and handheld ultrasound systems.Ultrasonography (Seoul, Korea) · 2026
- Energy consumption of standard and contrast-enhanced mammography: a step towards sustainable breast imagingEuropean Radiology · 2026
- Energy performance of MRI systems: on-site validation and comparison with manufacturer declarationsEuropean Radiology Experimental · 2026
- Towards Sustainable Magnetic Resonance Imaging: Insights from long-term, high-resolution energy recordings across an entire scanner fleetarXiv · 2026
- Monitoring building HVAC runtime using low-cost acoustic sensors.JASA express letters · 2026