Energy performance of MRI systems: on-site validation and comparison with manufacturer declarations
Andrea Roletto, Matteo Verga, Anna Savio, Gian Luca Viganò, Simone Zanoni
University of Brescia · Azienda Socio Sanitaria Territoriale degli Spedali Civili di Brescia
Key points
From the paper's conclusion- This study aims to evaluate the actual energy consumption of two generations of 1.5-T magnetic resonance imaging (MRI) scanners, quantify the benefits in terms of primary energy savings resulting from technological replacement, and compare field estimates of primary energy consumption with those reported in environmental product declarations (EPDs).
Abstract
OBJECTIVES: This study aims to evaluate the actual energy consumption of two generations of 1.5-T magnetic resonance imaging (MRI) scanners, quantify the benefits in terms of primary energy savings resulting from technological replacement, and compare field estimates of primary energy consumption with those reported in environmental product declarations (EPDs). MATERIALS AND METHODS: Two 1.5-T MRI scanner models, the old model version and its new model replacement, were monitored using a power quality analyzer connected to the electrical cabinet. Electrical power consumption data were collected over 2-week periods, both before and after the scanner replacement. Primary energy consumption was projected over 10 years, and the resulting values were compared with those reported in the EPDs for the two scanners. RESULTS: Over 10 years, cumulative energy consumption is estimated to be 1,010.4 MWh for the new unit versus 1,206.7 MWh for the old unit, corresponding to a 16.3% reduction. Considering the range of European primary energy factors (PEFs), energy savings varied from 235.6 to 687.1 MWh. Comparison with EPDs revealed significant discrepancies (± 40%) depending on the national PEF used, demonstrating that EPDs can both overestimate and underestimate actual energy consumption. CONCLUSION: Replacement of an old MRI model resulted in measurable energy savings, particularly in non-productive phases. However, EPDs do not always reflect clinical operation or the impact of national energy mixes. While energy efficiency is central to sustainable radiology, it should not be the sole driver for equipment replacement, which must remain primarily guided by clinical and diagnostic criteria. RELEVANCE STATEMENT: For a radiology department focused on more sustainable practices, it is essential to have accurate data on the environmental performance of medical imaging equipment, which should not be based solely on EPDs, but on real data based on usage patterns and national energy mixes. KEY POINTS: Replacing the old MRI scanner reduced energy consumption by 16.3%, mainly due to lower use in non-productive modes. Over a 10-year operational period, the primary energy consumption savings varied from 235.6 to 687.1 MWh. A discrepancy emerged between EPD-reported and real-world measurements, highlighting the importance of on-site validation for sustainability assessments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards Sustainable Magnetic Resonance Imaging: Insights from long-term, high-resolution energy recordings across an entire scanner fleetarXiv · 2026
- Data-Driven MRI Workflow Optimization and Platform Modernization: An Imaging Informatics and Energy Sustainability Study in a Resource-Limited Setting.Journal of imaging informatics in medicine · 2026
- Data-driven Assessment of Magnetic Resonance Imaging (MRI) Utilization and Aggregate Electrical Demand in U.S. Healthcare Facilities.Academic radiology · 2026
- Measuring the Environmental Impact of MRI and CT: A Life Cycle AssessmentJournal of the American College of Radiology · 2025
- Energy consumption of standard and contrast-enhanced mammography: a step towards sustainable breast imagingEuropean Radiology · 2026
- Environmental impact of a radiology department: greenhouse gas emissions and primary energy accounting and strategies for sustainable practice.Insights into imaging · 2026