Impact of acoustic property uncertainty on transcranial ultrasound simulation
J. Q. Kosciessa, T. D. Webb, E. Martin, B. Treeby, K. Butts Pauly, L. Verhagen, S. Pichardo
Key points
From the paper's conclusion- These results indicate that uncertainty in skull acoustic properties is a major source of variability in TUS exposure estimates, and that refined measurements of these properties, together with explicit reporting of exposure uncertainty, are needed to support safe, effective, and reproducible transcranial ultrasound planning.
Abstract
Subject-specific simulations of wave propagation are commonly used to plan and report acoustic and thermal exposure during transcranial ultrasound (TUS) interventions. Their accuracy depends on acoustic and thermal tissue property assumptions, including sound speed, density, and attenuation. Reported property measurements vary widely across the literature, precluding consensus point estimates for simulation. How this property uncertainty affects estimated exposure has not been quantified. We compile literature-based bounds for seven acoustic and thermal properties of skull and brain tissue and assess their individual impact on simulated intensity and temperature rise using two independent simulation platforms (PRESTUS/k-Wave and BabelBrain). Variations in skull attenuation and absorption produced the largest outcome uncertainty: peak intensity in the brain varied roughly threefold across the reported range of skull attenuation alone, while absorption most strongly affected temperature rise within the skull. The direction of each parameter's effect was consistent across platforms and, with one exception, across skull thickness. This allowed bounds to be combined into setups that bracket transmission and temperature rise. Compared to these extremes, default property sets of three simulation toolboxes (k-Plan, BabelBrain, PRESTUS) yielded similar brain intensity but divergent skull heating, driven mainly by differing assumptions on skull absorption and heat capacity. These results indicate that uncertainty in skull acoustic properties is a major source of variability in TUS exposure estimates, and that refined measurements of these properties, together with explicit reporting of exposure uncertainty, are needed to support safe, effective, and reproducible transcranial ultrasound planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental Validation of Skull Acoustic Modelling Strategies for Transcranial Focused Ultrasound Simulation: A Cross-Comparison Study.Physics in medicine and biology · 2026
- Computational Modeling of the Acoustic Radiation Force and Viscoelastic Displacement for Transcranial Focused Ultrasound NeuromodulationmedRxiv · 2026
- Calibrated simulations for dynamic focusing of ultrasound through the temporal window.Cell reports methods · 2026
- ITRUSST consensus on biophysical safety for transcranial ultrasound stimulationBrain stimulation · 2025
- Ultrasound Modulation of Brain Function.The Neuroscientist : a review journal bringing neurobiology, neurology and psychiatry · 2026
- Acoustic and Thermal Characterization of Tissue-Mimicking Phantoms for Modeling Focused Ultrasound Hyperthermia.IEEE transactions on ultrasonics · 2026