Experimental Validation of Skull Acoustic Modelling Strategies for Transcranial Focused Ultrasound Simulation: A Cross-Comparison Study.
Han Li, Aidan Horner, Zhihong Huang
School of Physics, Engineering and Technology, University of York, Heslington, York, York, York, YO10 5DD, United Kingdom of Great Britain and Northern Ireland. · Department of Psychology, University of York, Heslington, York, YO10 5DD, United Kingdom of Great Britain and Northern Ireland.
Key points
From the paper's conclusion- Paired specimen-level comparisons did not establish a consistent accuracy advantage for any strategy.
Significance
These results indicate that current skull modelling strategies may reproduce the gross intracranial beam pattern while still misestimating the delivered acoustic exposure, focal coverage and localization.
Abstract
Objective Accurate acoustic modelling of the skull is important for simulation-guided transcranial focused ultrasound (tFUS) applications. However, existing skull parameterization strategies vary widely in complexity and reported accuracy. This study systematically evaluates commonly used modelling approaches to determine their predictive performance across multiple frequencies and skull types. Approach Five skull modelling strategies were assessed: two voxel-wise linear mapping models, one with fixed attenuation coefficient and one with Hounsfield unit (HU)-dependent attenuation coefficients; two three-layer models with fixed and skull-density ratio (SDR) dependent attenuation coefficient; and one single-layer fixed-parameter model. Nineteen regions of interest from five historical and two Thiel-embalmed human skulls were examined at 220 kHz, 680 kHz and 1000 kHz, bowl-surface source fields were reconstructed using holography, and intracranial pressure distributions were captured with a needle hydrophone. Measurements were compared with k-Wave simulations. Results Average simulated peak pressure errors (ɛp) ranged from 20% to 31% across frequencies, whereas intensity errors (ɛI) reached 41% to 81%. Errors in -6 dB focal volume ranged from 18% to 65%, and focal position discrepancies were typically several millimeters. Simulation generally predicted smaller insertion losses than measured, indicating a tendency to underestimate skull-related attenuation and overestimate transmitted intracranial exposure. Among the five modelling strategies, the model with the lowest observed error varied with frequency and outcome. Paired specimen-level comparisons did not establish a consistent accuracy advantage for any strategy. Significance These results indicate that current skull modelling strategies may reproduce the gross intracranial beam pattern while still misestimating the delivered acoustic exposure, focal coverage and localization. Such errors are directly relevant to tFUS dose selection, target coverage and safety-margin interpretation, underscoring the need for refined parameterization methods for reliable tFUS treatment planning.
.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of acoustic property uncertainty on transcranial ultrasound simulationbioRxiv · 2026
- Adaptive Positioning and Targeting Optimization in Transcranial Focused Ultrasound (tFUS): A Numerical Study.Ultrasound in medicine & biology · 2026
- Calibrated simulations for dynamic focusing of ultrasound through the temporal window.Cell reports methods · 2026
- Computational Modeling of the Acoustic Radiation Force and Viscoelastic Displacement for Transcranial Focused Ultrasound NeuromodulationmedRxiv · 2026
- A deep‐learning model for one‐shot transcranial ultrasound simulation and phase aberration correctionMedical Physics · 2025
- Effects of sparse array geometry and element arrangement on low-intensity transcranial focused ultrasound.PloS one · 2026