World journal of emergency medicine

Utilization of hands-free carotid artery Doppler ultrasound for noninvasive central venous pressure measurement: a prospective observational study.

Zhao Menglong, Mei Qing, Zhang Lin, Zhang Lei, Fan Xiaoqin, Tong Fei, Chen Jian, Zhou Min

Department of Intensive Care Unit, the First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei 230001, China.

Observational · n=94

Key points

From the paper's conclusion
  • A univariate linear regression equation, CVP = 6.703 × (H-cVTI/H-jVTI) + 3.603, yielded an area under the ROC curve (AUC) of 0.961 (95%CI: 0.926-0.996) for predicting CVP > 10 cmH2O, with an optimal cutoff of 1.12, sensitivity of 90.5%, and specificity of 93.5%.

Abstract

BACKGROUND: Conventional central venous pressure (CVP) measurement requires central venous catheterization, which can be difficult to perform promptly in emergency settings. Bedside ultrasound provides an alternative approach by extracting CVP-related parameters. This study aims to evaluate the feasibility and accuracy of a hands-free carotid artery Doppler ultrasound (hCDU) method for noninvasive estimation of CVP in mechanically ventilated patients. METHODS: This prospective study enrolled adult patients admitted to the Intensive Care Unit and receiving invasive mechanical ventilation from November 2024 to February 2025. Within 24 h of admission, blood pressure, heart rate, and CVP were collected. Conventional ultrasound was used to obtain carotid and internal jugular velocity time integrals (C-cVTI and C-jVTI) and the cross-sectional areas of the carotid artery (CSAa) and internal jugular vein (CSAj). Immediately, the carotid artery velocity time integral (H-cVTI) and the internal jugular vein velocity time integral (H-jVTI) were obtained by using the hCDU. Inter-rater reliability was assessed, and agreement between conventional ultrasound and hCDU measurements was evaluated using Bland-Altman plots. Correlation analyses and receiver operating characteristic (ROC) curve analysis were performed to evaluate correlations and diagnostic efficacy. RESULTS: A total of 94 patients were included. CSAj (r = 0.403, P<0.001) and CSAj/CSAa (r = 0.421, P<0.001) showed moderate positive correlations with CVP, whereas CSAa showed no significant correlation. H-cVTI (r = 0.384, P<0.001) was weakly positively correlated with CVP, H-jVTI (r = -0.677, P<0.001) was moderately negatively correlated, and H-cVTI/H-jVTI (r = 0.812, P<0.001) strongly positively correlated. A univariate linear regression equation, CVP = 6.703 × (H-cVTI/H-jVTI) + 3.603, yielded an area under the ROC curve (AUC) of 0.961 (95%CI: 0.926-0.996) for predicting CVP > 10 cmH2O, with an optimal cutoff of 1.12, sensitivity of 90.5%, and specificity of 93.5%. CONCLUSION: The H-cVTI/H-jVTI ratio obtained via the hCDU method shows strong correlation with invasively measured CVP and may represent an alternative technique for noninvasive CVP estimation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.