AI-Enabled Wearable Ultrasound for Noninvasive Central Venous Pressure Monitoring.
Liping Dong, Xingxuan Zhang, Meng Li, Yi Li, Jingyi Guo, Shaorong Lu, Zhenyu Peng, Chenxi Zheng,
Department of Ultrasound, Shanghai Sixth People's Hospital Affiliated to Shanghai Jiao Tong University School of Medicine; Shanghai Institute of Ultrasound in Medicine, Shanghai 200233, China. · Department of Clinical Research Center, Shanghai Sixth People's Hospital, Shanghai Jiao Tong University of Medicine, Shanghai 200233, China.
Key points
From the paper's conclusion- These findings indicate that an AI-integrated wearable ultrasound device can provide accurate, noninvasive bedside assessment of CVP and may offer a promising alternative to invasive catheters for hemodynamic monitoring in acute and critical care settings.
Abstract
Central venous pressure (CVP) is a key indicator of right ventricular preload, but measurement with a central venous catheter (CVC) is invasive, time-consuming, and unsuitable for bedside monitoring. We developed and clinically evaluated an artificial intelligence (AI)-enabled wearable ultrasound device for noninvasive estimation of elevated CVP by imaging the internal jugular vein (IJV) and common carotid artery (CCA) in acute and critical patients. In this prospective multi-center clinical study, 349 patients admitted to intensive care units (ICUs) at 2 tertiary hospitals underwent neck vascular imaging with the wearable ultrasound device, while CVP was measured simultaneously via CVC as the reference standard. The dual-decoder spatiotemporal attention network (DSTA-Net) segmentation model automatically quantified IJV and CCA cross-sectional areas, and the dual-modality multilayer perceptron (DM-MLP) integrated these vascular indices with basic clinical parameters to predict elevated CVP (≥8 mmHg). DSTA-Net showed excellent agreement and strong correlation with expert manual measurements while achieving superior segmentation accuracy compared with the baseline segmentation model. Meanwhile, the proposed DM-MLP improved prediction performance by 4% to 8% over conventional baseline architectures, achieving AUCs of 0.91 (internal test set) and 0.87 (external test set). These findings indicate that an AI-integrated wearable ultrasound device can provide accurate, noninvasive bedside assessment of CVP and may offer a promising alternative to invasive catheters for hemodynamic monitoring in acute and critical care settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Machine Learning-Based Identification of Patients with Elevated Central Venous Pressure Using Features Extracted from Photoplethysmography and Arterial Blood Pressure Waveforms.Research square · 2026
- Validation of a Novel Point-of-Care Ultrasound Method for Estimating Central Venous Pressure in Critically Ill Patients.Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine · 2026
- Utilization of hands-free carotid artery Doppler ultrasound for noninvasive central venous pressure measurement: a prospective observational study.World journal of emergency medicine · 2026
- Pediatric Venous Excess Ultrasound Score (P-VExUS): a novel approach to assess central venous pressure in the PICU.European journal of pediatrics · 2026
- Edge-Native Non-Contact Heart Rate Monitoring via Wi-Fi CSI on Low-Cost MicrocontrollersACM Transactions on Embedded Computing Systems · 2026
- Continuous Cuffless Blood Pressure Estimation Using a Multimodal Conv1D-LSTM Architecture and the SOUNDI® Wearable Device.Biosensors · 2026