Bailout Guiding Catheter Exchange Using Medilizer: A Case Report.
Yoshitaka Yamaguchi, Kei Miyata, Daisuke Shimbo, Akira Yoshida, Tatsuro Takada
Department of Cerebrovascular Medicine, Teine Keijinkai Hospital, Sapporo, Hokkaido, Japan. · Department of Neurosurgery, Teine Keijinkai Hospital, Sapporo, Hokkaido, Japan.
Key points
From the paper's conclusion- This study provides preliminary evidence that intraoperative AI notifications may reflect differences in technical behavior according to operator certification status. Although the adjusted association was not statistically significant, these findings suggest the potential of AI-assisted systems as proof-of-concept tools for evaluating technical proficiency. Further validation involving a larger number of operators is required.
Abstract
OBJECTIVE: Medilizer (Medikit, Tokyo, Japan) is a distal anchoring exchange wire equipped with a soft, flexible self-expanding stent at the distal tip, designed to facilitate exchange maneuvers and device delivery. We report the use of the Medilizer for bailout guiding catheter exchange while preserving distal access. CASE PRESENTATION: A 72-year-old man presented with progressive cerebral ischemia caused by acute right internal carotid artery (ICA) occlusion with contralateral chronic ICA occlusion. Percutaneous transluminal angioplasty (PTA) and Wingspan stent (Stryker Neurovascular, Fremont, CA, USA) placement in the right ICA were attempted under proximal balloon protection using a balloon-guiding catheter. After lesion crossing, the Medilizer was deployed from the proximal M2 to the proximal M1 segment. Cerebral angiography revealed severe cavernous ICA stenoses and multiple thrombi in the petrous ICA. During attempted PTA under proximal balloon protection, rupture of the balloon-guiding catheter necessitated an unexpected guiding catheter exchange. Because exchanging the entire system risked the loss of distal access, guiding catheter exchange was successfully attempted while maintaining the Medilizer in place using stepwise coaxial catheter advancement. Subsequent angioplasty and Wingspan stent placement achieved adequate revascularization without complications. CONCLUSION: Medilizer may provide an effective bailout option for unexpected guiding catheter exchange after lesion crossing by preserving distal access during neuroendovascular procedures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Successful Endovascular Retrieval of a Partially Disrupted Catheter Using a Crossing Catheter: A Bailout Technique for Severely Calcified Peripheral Lesions.Annals of vascular diseases · 2026
- Intravascular Lithotripsy Balloon Entrapped in Right Coronary Artery Requiring Surgical Bailout.JACC. Case reports · 2026
- A steerable microcatheter-stabilizer strategy for stent-assisted coiling of a wide-neck middle cerebral artery aneurysm with an acutely angulated incorporated branch.Surgical neurology international · 2026
- Guide-sheath-assisted bailout technique for intra-procedural single-leaflet device attachment using the PASCAL system: a case report.European heart journal. Case reports · 2026
- Transcatheter Mitral Valve-in-Valve Replacement Complicated by Valve Embolization: Percutaneous Rescue With a Double-Rail Technique.JACC. Case reports · 2026
- When the RotaWire Crosses but Cannot Advance: Dual-Lumen Microcatheter-Assisted RotaWire Delivery.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026