Comparative dosimetric analysis of flattening-filter and flattening-filter-free photon beams in radiotherapy based on TRS-398 calibration protocol.
Sakinah Sakinah, Primasari Cahya Wardhani, Arie Chintya Martania, Wahyu Dwi Lestari
Department of Physics, Faculty of Engineering and Science, Universitas Pembangunan Nasional "Veteran" Jawa Timur, Indonesia. · Calibration, Radiation Measurement, and Radiotherapy Instrument Installation, Balai Pengamanan Alat dan Fasilitas Kesehatan, Surabaya, Indonesia.
Key points
From the paper's conclusion- Both FF and FFF photon beams meet the TRS-398 standards for how accurate the radiation dose should be. But, the FFF mode is more stable, so it could be used more in hospitals, especially for very precise radiotherapy. These results show that FFF technology can make radiotherapy better, especially in places where the newest equipment is not available.
Abstract
BACKGROUND: In external beam radiotherapy, delivering an accurate radiation dose is essential to ensure treatment effectiveness while minimising harm to surrounding healthy tissue. In Indonesia, the adoption of flattening-filter-free (FFF) technology remains limited, making its evaluation highly relevant for supporting more equitable radiotherapy services nationwide. The FFF technique offers advantages such as higher dose rates, shorter treatment time, and reduced head scatter, providing opportunities for faster and more precise treatment. This study compares the output dose deviations between flattening-filter (FF) and FFF modes using the IAEA TRS-398 dosimetry protocol. MATERIALS AND METHODS: Measurements were performed using a Varian TrueBeam linear accelerator with 6 MV and 10 MV FF beams and a 6 MV FFF beam. Absolute dose calibration was conducted with a calibrated ionisation chamber and electrometer in a water phantom at a reference depth of 10 cm. Data were collected from April to August 2025, with five repeated measurements for each energy. RESULTS: The results showed dose deviations of 0.908% (6 MV), 0.876% (10 MV), and 0.020% (6 FFF), which are all within the ±2% allowed by TRS-398. Among the evaluated modes, the FFF beam demonstrated the highest output stability. CONCLUSIONS: Both FF and FFF photon beams meet the TRS-398 standards for how accurate the radiation dose should be. But, the FFF mode is more stable, so it could be used more in hospitals, especially for very precise radiotherapy. These results show that FFF technology can make radiotherapy better, especially in places where the newest equipment is not available.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dosimetric comparison of 6-MV flattened and flattening filter-free photon beams using Monte Carlo simulationNuclear Engineering and Technology · 2026
- Monte Carlo evaluation of small-field output correction factors of a microDiamond detector in flattened and flattening-filter-free photon beams for stereotactic radiotherapy.Radiation and environmental biophysics · 2026
- ACPSEM guideline regarding the adoption of the IAEA TRS-398 Code of Practice Revision 1.Physical and engineering sciences in medicine · 2026
- Pulse-width-specific phase space modeling and reference beam characterization for UHDR electron LINAC in FLASH-RTPhysics in Medicine and Biology · 2026
- Dosimetric Characterization and Workflow Optimization of the FLASH-SARRP for Reliable Preclinical Radiobiological StudiesbioRxiv · 2026
- Beam data and model validation for Elekta SRS cone collimator in 6MV and 6MVFFF beams.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026