ACPSEM guideline regarding the adoption of the IAEA TRS-398 Code of Practice Revision 1.
Joerg Lehmann, Jeffrey Barber, Andrew Alves, Simon Berke, Rhonda Brown, Peter Harty, Brendan Healy, Robin Hill,
Institute of Medical Physics, University of Sydney, Sydney, NSW, Australia. Joerg.Lehmann@sydney.edu.au. · Institute of Medical Physics, University of Sydney, Sydney, NSW, Australia.
Key points
From the paper's conclusion- For specialised FFF machines such as CyberKnife® and TomoTherapy®, and for Gamma Knife®, the ACPSEM recommends the use of IAEA TRS-483 Code of Practice.
Abstract
The IAEA Technical Report Series 398 Code of Practice for the determination of absorbed dose in external beam radiotherapy was endorsed at the Engineers and Physical Scientists in Medicine (EPSM) conference in 2001 and has been widely adopted in Australia and New Zealand. This guideline provides recommendations from the Australasian College of Physical Scientists and Engineers in Medicine (ACPSEM) following the recent publication of TRS-398 Revision 1. It is jointly authored by the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA), the Reference Dosimetry Advisory Group (RDAG) and the ACPSEM Radiation Oncology Specialty Group (ROSG). In addition to recommendations, the guideline discusses the implications of TRS-398 Revision 1 for the dosimetry of megavoltage photons, electrons and protons. TRS-398 Revision 1 is recommended for adoption in accordance with the recommendations provided in this guideline, including for FFF beams. For specialised FFF machines such as CyberKnife® and TomoTherapy®, and for Gamma Knife®, the ACPSEM recommends the use of IAEA TRS-483 Code of Practice. MR linacs are covered by the ACPSEM position paper: dosimetry for magnetic resonance imaging linear accelerators and in AAPM TG-351. For kilovoltage photons, the recommendation to use the AAPM TG-61 protocol for radiotherapy sources is retained and now specifically also includes cabinet irradiators used in radiobiology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative dosimetric analysis of flattening-filter and flattening-filter-free photon beams in radiotherapy based on TRS-398 calibration protocol.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2026
- Harmonizing spatially fractionated radiation therapy: consensus and roadmap from an ESTRO workshop.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Ion recombination correction in reference dosimetry for pencil beam scanned proton beamsMedical Physics · 2026
- Adoption and Practice Patterns of TG-263 Standardized Nomenclature for Breast Radiation Therapy in a Statewide Quality Consortium.Practical radiation oncology · 2026
- Development of a Statistical Framework for Optimizing Image-Guidance Frequency in Whole-Breast Radiotherapy.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Cumulative dose evaluation in clinical reirradiation – Consensus guidance on technical considerations by the ESTRO reirradiation focus groupRadiotherapy and Oncology · 2025