Prospective controlled ex vivo observational study on changes in elasticity of ureteral stents following clinical use.
Marina Hoffmann, Benedikt Ebner, Nikolaos Pyrgidis, Iulia Blajan, Troya Georgieva, Marc Kidess, Moritz Happe, Yannic Volz,
Department of Urology, LMU Klinikum, Munich, Germany.
Key points
From the paper's conclusion- Ureteral stents incorporating HybridPRO Technology demonstrated significantly greater elongation under standardized load after use compared to a control stent. These findings indicate material changes ex vivo; however, the clinical implications regarding comfort or quality of life remain to be determined in future studies.
Abstract
INTRODUCTION: Ureteral stents significantly impair health-related quality of life. Advances in stent material, particularly in polyurethane compositions, may mitigate adverse effects. This study evaluated ex vivo changes in the elasticity of ureteral stents made from a novel thermoplastic polyurethane composite (HybridPRO Technology by Urotech) compared to a control stent. MATERIAL AND METHODS: In this prospective controlled ex vivo observational study, 97 patients undergoing ureteroscopy received one of two types of tethered 7 Fr, 28 cm ureteral stents. Eleven patients underwent bilateral procedures, yielding a total of 108 stents (54 per group). Group 1 received a polyurethane stent with HybridPRO Technology and group 2 received a polyurethane stent without HybridPRO Technology. After removal, stents were measured at the patient's bedside using a hanging apparatus with a predetermined 400 g load. In vitro (room temperature) and ex vivo elongations were recorded, and the net difference was calculated. RESULTS: In vitro measurements showed a median elongation of 4 mm (IQR: 4-4) in group 1 vs 2 mm (IQR: 2-2) in group 2. Ex vivo, group 1 elongated significantly more (median, 12 mm; IQR: 11-13) compared to group 2 (median: 5 mm; IQR: 4-6; p <0.001). The net difference was larger in group 1 (median: 10 mm; IQR, 9-11) compared to group 2 (median: 3 mm; IQR: 2-4; p <0.001). CONCLUSIONS: Ureteral stents incorporating HybridPRO Technology demonstrated significantly greater elongation under standardized load after use compared to a control stent. These findings indicate material changes ex vivo; however, the clinical implications regarding comfort or quality of life remain to be determined in future studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Choosing the optimal ureteral stent design, size, and material in different situations: a review from the European association of urology endourology.World journal of urology · 2026
- Effects of segment chemistry on mechanical performance and hemocompatibility in polyurethane heart valve scaffolds.Journal of materials chemistry. B · 2026
- A novel ureteral stent to reduce urinary reflux: in vitro and ex vivo testing.International urology and nephrology · 2026
- A Comparative In Vivo Study of Small-Diameter Vascular Grafts Fabricated from Electrospun Tecothane and Tecoflex Polyurethanes.ACS biomaterials science & engineering · 2026
- Manometry-based comparison of ex-vivo versus in-vivo inflation profiles among antibiotic-coated inflatable penile prostheses and the impact of progressive modeling in instances of concomitant Peyronie's disease.Urology · 2026
- Enhanced biocompatibility and efficacy of a newly developed plastic biliary stent: a porcine model study.BMC gastroenterology · 2026