Urology

Manometry-based comparison of ex-vivo versus in-vivo inflation profiles among antibiotic-coated inflatable penile prostheses and the impact of progressive modeling in instances of concomitant Peyronie's disease.

Briana Goddard, Sarah Parker, Ryan Terlecki

Wake Forest University School of Medicine, Department of Urology; Winston Salem, North Carolina, United States. Electronic address: brianalgoddard@gmail.com. · Wake Forest University School of Medicine, Department of Urology; Winston Salem, North Carolina, United States.

Observational · n=50

Key points

From the paper's conclusion
  • Higher cylinder pressures were achieved in-vivo despite fewer pumps and lower fill volumes. Plaque disruption with modeling allowed for greater accommodation of fluid within the cylinders. The identified differences between benchtop and intracorporal performance point to the relevance of the host tissue toward device performance.

Abstract

OBJECTIVES: To better understand how the biomechanical properties of penile tissues impact inflatable penile prosthesis (IPP) performance, we sought to compare device metrics both ex- and in-vivo. IPPs allow rigidity for intercourse by internal fluid pressure, but there is currently little known about how device performance changes after being placed intracorporeal. We assessed the available array of IPPs with standardized antibiotic coatings (AMS 700™, Boston Scientific) using fluid manometry to compare ex- and in-vivo device properties. METHODS: 50 patients receiving an IPP were prospectively assessed. For both ex- and in-vivo measurements, the number of pumps and fluid volume to inflate the IPP to full rigidity were counted. Fluid manometry was used to measure internal device pressure. For patients with demonstrable curvature, both pre-and post- modeling parameters were assessed. RESULTS: Full rigidity in-vivo was achieved with fewer pumps (25.8 vs 29.7 pumps, p<0.001) and lower fluid volume (43.6 vs 65.8 mL, p <0.001), yet significantly higher pressure (15.5 vs 14.8 psi, p=0.015) than ex-vivo measurements. Modeling produced a significant increase in pumps to full rigidity (20.5 vs 28.1; p <0.001), fill volume (36.1 vs 43.3; p <0.001) and pressure (13.7 vs 16.9 psi; p < 0.001). CONCLUSION: Higher cylinder pressures were achieved in-vivo despite fewer pumps and lower fill volumes. Plaque disruption with modeling allowed for greater accommodation of fluid within the cylinders. The identified differences between benchtop and intracorporal performance point to the relevance of the host tissue toward device performance.

Read the paper