International urology and nephrology

A novel ureteral stent to reduce urinary reflux: in vitro and ex vivo testing.

Manikantam Gaddam, Ashley Phord-Toy, Suhaansai Thatikonda, John Jay Ogunkeye, Taylor Hughes, Rachelle Balili Kanegawa, Pranaya Venkatapuram, Emily Chang,

Department of Urology, Stanford University School of Medicine, Palo Alto, CA, USA. · Division of Urology, Cook Children's Medical Center, Fort Worth, TX, USA.

Preclinical

Key points

From the paper's conclusion
  • This novel ureteral stent reduced retrograde flow, while preserving antegrade drainage through a pressure-responsive distal retention element under benchtop conditions.

Abstract

PURPOSE: While traditional double-J ureteral stents reliably relieve obstruction, they often cause discomfort and pain due to vesicoureteral reflux resulting from disruption of the ureter's natural anti-refluxing mechanism. Through in vitro hydrodynamic testing and ex vivo extraction studies, this study evaluates a novel ureteral stent designed to address reflux. METHODS: This novel stent consists of 1) a flexible polymer conduit, 2) a collapsible distal segment functioning as a one-way valve to promote antegrade urine flow, and 3) a flexible conical distal retention element to reduce retrograde flow. In vitro flow rate testing consisted of two cavities connected by a 5 mm conduit simulating the ureter. Pressure differentials were set for antegrade flow (renal pelvis  =  15 cm H2O, bladder = 0 cm H2O) and retrograde flow (renal pelvis = 0 cm H2O, bladder  = 50 cm H2O). Ten ex vivo porcine kidneys were stented with a 6 Fr novel stent or a commercially available 6 Fr JJ stent (BARD Inlay Optima™). Stents were positioned with the proximal end in the renal pelvis and distal end at the ureterovesical junction, then connected to a digital force gauge for extraction testing. RESULTS: In vitro flow testing demonstrated that the novel stent preserved antegrade drainage, while reducing retrograde flow. During axial traction of ex vivo testing, the conical distal element consistently deformed outward, reducing radial wall contact and avoiding distal wedging. CONCLUSIONS: This novel ureteral stent reduced retrograde flow, while preserving antegrade drainage through a pressure-responsive distal retention element under benchtop conditions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A novel ureteral stent to reduce urinary reflux: in vitro and ex vivo testing — Science Skim