Journal of contemporary brachytherapy

Dosimetric impact of computed tomography slice thickness-induced applicator reconstruction uncertainties in three-dimensional brachytherapy for cervical cancer.

Wei Huang, Haixia Cui, Dongshan Wei, Xiuquan Li, Qiong Wang, Qingfeng Jiang

Department of Oncology, Laboratory of Immunity, Inflammation & Cancer, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China. · Chongqing Engineering Research Center of Pharmaceutical Sciences, Chongqing Medical and Pharmaceutical College, Chongqing, China.

Observational

Key points

From the paper's conclusion
  • CT slice thickness ≥ 2.50 mm introduces significant reconstruction inaccuracies, leading to clinically relevant dose deviations for HR-CTV and rectum. Reconstructing both 0.625 mm (for applicator reconstruction) and 2.50 mm (for contouring) images from a single scan is recommended for high-density applicator brachytherapy.

Abstract

PURPOSE: To evaluate the impact of computed tomography (CT) slice thickness on reconstruction accuracy of high-density tandem-ovoid applicators and resulting dose deviations in cervical cancer brachytherapy. MATERIAL AND METHODS: Seventeen cervical cancer patients treated with CT-based brachytherapy using stainless steel tandem-ovoid applicators (2.50 mm slice thickness) were retrospectively analyzed. Homologous 0.625 mm, 1.25 mm, and 5.00 mm image sets were reconstructed from original CT raw data. Applicators were reconstructed on each thickness, and treatment plans from 2.50 mm images were copied to other sets. Applicator displacements, reconstruction uncertainties, and dose deviations at 1.25 mm, 2.50 mm, and 5.00 mm were compared with a 0.625 mm reference. RESULTS: Slice thickness-induced partial volume effect caused applicator reconstruction uncertainties, predominantly in superior-inferior ovoid displacements, with maximum uncertainties approaching slice thickness. Each 1 mm superior-inferior displacement caused high-risk clinical target volume (HR-CTV) D90 deviations of 1.23% (right) and 2.47% (left). Each 1 mm anterior-posterior displacement resulted in rectal D2cc deviations of 1.71% (right) and 2.34% (left). Thicker slices (2.50 and 5.0 mm) led to average HR-CTV D90 overestimations of 2.86% and 11.11% (maximum 7.06% and 25.55%). Rectal D2cc was underestimated by an average of 1.40% and 4.43% (maximum: 6.67% and 8.39%). CONCLUSIONS: CT slice thickness ≥ 2.50 mm introduces significant reconstruction inaccuracies, leading to clinically relevant dose deviations for HR-CTV and rectum. Reconstructing both 0.625 mm (for applicator reconstruction) and 2.50 mm (for contouring) images from a single scan is recommended for high-density applicator brachytherapy.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers