Geometric accuracy of applicator and needle reconstruction in gynecological brachytherapy using low-field magnetic resonance imaging.
Tibor Major, Júlia Vizkeleti, Zoltán Takácsi-Nagy
Radiotherapy Centre, National Institute of Oncology, Budapest, Hungary.
Key points
From the paper's conclusion- The 0.55T MR scanner can provide clinically acceptable accuracy in the reconstruction of brachytherapy (BT) applicators and needles. MR markers allow greater precision in determining the tips of applicators. The accuracy of reconstruction using MR images is comparable to that achieved with CT images.
Abstract
PURPOSE: To evaluate the geometric accuracy of applicator reconstruction in gynecological brachytherapy using low-field magnetic resonance imaging. MATERIAL AND METHODS: A 0.55T magnetic resonance (MR) scanner was used in this study. First, accuracy of needle reconstruction based on MR and computed tomography (CT) images were compared. Seven plastic needles were inserted into three phantoms, which were scanned with MR and CT. Five observers determined tip and entry points of the needles on both types of images. Needle lengths and tip coordinates were compared. Then, inter-observer variation in localization of applicator and needle tips in patients was evaluated on MR images. Twenty fractions of eleven patients treated with Venezia applicator with needles were selected. In tandem and ovoid tubes, MR markers were inserted, with only air in the needles. Differences between observers in determining the coordinates were depicted using standard deviation (SD). RESULTS: The mean difference between needle depths was 1.16 ±0.8 mm measured on MR and CT images. Uncertainties among observers in determining the position of tips were generally small, with SD of less than 1.0 mm for both imaging modalities. A good correlation (R 2 > 0.99) was found between the tip coordinates calculated on MR and CT images. In patients, SD of applicator tips with MR markers was smaller than that of needle tips without markers (X: 0.18 vs. 0.27 mm, Y: 0.24 vs. 0.58 mm, Z: 0.21 vs. 0.34 mm). Regarding direction, the greatest variability was constantly observed in the Y direction that corresponded to the greater slice thickness. CONCLUSIONS: The 0.55T MR scanner can provide clinically acceptable accuracy in the reconstruction of brachytherapy (BT) applicators and needles. MR markers allow greater precision in determining the tips of applicators. The accuracy of reconstruction using MR images is comparable to that achieved with CT images.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Three-dimensional MRI-based FXO Gel Dosimetry for Verification of HDR Venezia Applicator Cervix BrachytherapyJournal of Physics: Conference Series · 2026
- Prospective Paired Comparison of MRI- and CT-Based Planning for Adjuvant Vaginal Cuff Brachytherapy in Endometrial Cancer: An Exploratory StudyRadiation · 2026
- Characterizing magnetic field coupling in MRI-guided radiotherapy by solving a Dirichlet problem with sparsely sampled boundary conditions.Physics in medicine and biology · 2026
- Comparative effectiveness and safety of magnetic resonance imaging-guided adaptive brachytherapy versus conventional brachytherapy in locally advanced cervical cancer: a systematic review and meta-analysisFrontiers in Oncology · 2026
- Dose distribution in a cervix phantom quality-assurance using two computer modeling approaches.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Magnetic Resonance Imaging Compared With Computed Tomography for Segmentation-Based Cavity Volume Measurements in Gelatin Head Phantoms.The American journal of forensic medicine and pathology · 2026