Journal of Physics: Conference Series

Three-dimensional MRI-based FXO Gel Dosimetry for Verification of HDR Venezia Applicator Cervix Brachytherapy

K M Alexander, JD Jutras, G S Lee, C Fallone, W R Gill, T Meyer

Key points

From the paper's conclusion
  • This work applies 3D MRI-based FXO gel dosimetry as an end-to-end verification method for HDR cervix brachytherapy using the Elekta Venezia applicator in both intracavitary and hybrid interstitial geometries.

Abstract

This work applies 3D MRI-based FXO gel dosimetry as an end-to-end verification method for HDR cervix brachytherapy using the Elekta Venezia applicator in both intracavitary and hybrid interstitial geometries. Custom gel phantoms containing clinically relevant Venezia configurations were irradiated and read out using MRI. Measured dose distributions showed strong agreement with TG-43 treatment planning system calculations with 3D gamma pass rates exceeding 97% using 2%/2mm criteria. Residual disagreement was generally localized near the tandem and interstitial needles, where steep dose gradients, applicator-related geometric uncertainty, and non-linear FXO response limit quantitative interpretation. These results support MRI-based FXO gel dosimetry as a practical tool for commissioning and verification of complex gynaecological HDR brachytherapy workflows, with a clear pathway toward clinical error-detection applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers