Digital Thoracic Drainage System-Derived Circuit Pressure and Air-Flow Changes in Relation to Echocardiographic Indices of Cardiac Function after Chest Tube Placement.
Atilla Can, Tamer Altınok, Mehmet Burhan Oflaz, Abdullah İçli
Department of Thoracic Surgery, Selçuk University Faculty of Medicine, Konya, Türkiye. · Department of Thoracic Surgery, Necmettin Erbakan University Faculty of Medicine, Konya, Türkiye.
Key points
From the paper's conclusion- Measured circuit pressure appeared to reflect regulated drainage-system pressure more closely than pleural physiology. Most reduction in air leak occurred before the first echocardiographic assessment, potentially limiting the detection of cardiopulmonary associations. These findings support a cautious interpretation of digital drainage parameters and may help define the assessment window for future studies.
Abstract
OBJECTIVE: Chest tube drainage restores negative intrathoracic pressure, which might alter cardiac loading. Digital thoracic drainage systems continuously record pressure and airflow. We examined device-derived pressure and air-flow trends over 48 hours and echocardiographic changes between 24 and 48 hours. MATERIAL AND METHODS: Forty-one chest tube drainage episodes in 39 adult, non-mechanically ventilated patients were studied prospectively. Device records were analysed at 10-minute intervals, providing preset suction level, measured circuit pressure, and air flow. Echocardiography was performed at 24 and 48 hours. Continuous time-series analyses included 29 episodes, 28 of which were evaluable within the predefined 48-hour window. False discovery rate correction was applied to multiple comparisons. RESULTS: The preset suction level remained unchanged during the first 48 hours in 21 of 29 episodes (72%), and the measured circuit pressure was closely related to it at each time point (R2 = 0.838-0.955). Air leak decreased from a median of 61.6 mL/min at baseline to 2.3 mL/min at 24 hours (P < 0.001), with no further change at 48 hours (1.3 mL/min; P = 0.55). After correction, ejection fraction (q = 0.011) and mitral E velocity (q = 0.007) were higher at 48 hours. No device-derived variable was associated with echocardiographic changes after correction for multiple comparisons. CONCLUSION: Measured circuit pressure appeared to reflect regulated drainage-system pressure more closely than pleural physiology. Most reduction in air leak occurred before the first echocardiographic assessment, potentially limiting the detection of cardiopulmonary associations. These findings support a cautious interpretation of digital drainage parameters and may help define the assessment window for future studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Pressure Behavior of a Digital Drainage System.Innovations (Philadelphia, Pa.) · 2026
- Positional and Temporal Changes in Endotracheal Tube Cuff Pressure in Mechanically Ventilated Intensive Care Unit Patients: A Prospective Observational Study.Medical science monitor : international medical journal of experimental and clinical research · 2026
- Serial transthoracic echocardiography for hemodynamic monitoring and early warning of hypotension during elective cesarean delivery: a prospective observational study.Journal of clinical monitoring and computing · 2026
- Dissociation between airway and transpulmonary pressures during external chest wall restriction: a prospective self-controlled physiological study.Journal of thoracic disease · 2026
- Transthoracic echocardiography as a predictor of spinal anesthesia-induced hypotension in cesarean section: a prospective observational study.BMC anesthesiology · 2026
- Cardiac output and stroke volume dissociate during fluid loading in conscious sheep.American journal of physiology. Heart and circulatory physiology · 2026