Innovations (Philadelphia, Pa.)

The Pressure Behavior of a Digital Drainage System.

Alberdina W Klomp, Michiel Wagenaar, Ron G H Speekenbrink, Frans H C de Jongh

Department of Pulmonary Medicine, Medisch Spectrum Twente, Enschede, The Netherlands. · Department of Cardiothoracic Surgery, Thorax Centrum Twente, Medisch Spectrum Twente, Enschede, The Netherlands.

Key points

From the paper's conclusion
  • The Thopaz+ digital drainage system generates transient overshoot in negative intrapleural pressures immediately after activation. The clinical relevance of this phenomenon requires further investigation.

Abstract

OBJECTIVE: Digital chest drainage systems are widely used since they provide objective monitoring of air leak and may shorten drainage duration and hospital stay. However, the pressure behavior of these systems is unknown. This study aimed to evaluate whether the preset suction pressure of the Thopaz+ digital drainage system (Medela AG, Baar, Switzerland) corresponds to the actual measured intrapleural pressure during initial activation and also compares it with an analog water-seal system. METHODS: An in vitro model simulating the pleural space was developed using a ventilator, a test lung, and a sealed barrel containing a balloon to mimic lung expansion. Two configurations were tested: spontaneous breathing and mechanical ventilation. Intrapleural pressure was measured while the Thopaz+ system was set to -8 cm H2O suction. Additional experiments simulated a pneumothorax with adjustable air leaks (150 and 550 mL/min). Also, a 2-step protocol (-5 cm H2O followed by -8 cm H2O) was evaluated. Results were also compared with an analog water-seal drainage system. RESULTS: In both spontaneous and ventilated models, the Thopaz+ system showed a transient pressure overshoot during the first minutes. With a preset suction of -8 cm H2O, the measured intrapleural pressure reached -27 cm H2O. This negative overshoot decreased with increasing air leak. A gradual 2-step suction approach prevented pressures from dropping below -8 cm H2O. In contrast, the analog drainage system did not generate pressures below the preset level. CONCLUSIONS: The Thopaz+ digital drainage system generates transient overshoot in negative intrapleural pressures immediately after activation. The clinical relevance of this phenomenon requires further investigation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers