Development and validation of a standardized causality assessment tool for adverse events associated with medical devices.
Jeesa George, Adusumilli Pramod Kumar, Vivekanandan Kalaiselvan, Vibha Shetty, Anjana S Nair, Ananya Chakraborty, M Kiran, M G Sindu,
Department of Pharmacy Practice, Faculty of Pharmacy, M. S. Ramaiah University of Applied Sciences, Bengaluru, Karnataka, India. · Materiovigilance Program of India, Indian Pharmacopoeia Commission, Ministry of Health and Family Welfare, Government of India, Ghaziabad, Uttar Pradesh, India.
Key points
From the paper's conclusion- The validated causality assessment tool enhances consistency and reduces subjectivity among evaluators. It provides a reliable scoring system to support regulatory decisions, improve safety, and enable postmarket surveillance. This tool enables systematic assessment of device-related AEs across healthcare settings.
Abstract
BACKGROUND: Medical devices are crucial in health care; however, adverse events (AEs) require safety evaluation. Causality assessment determines the contribution of the device to AEs and guides decisions. India lacks a standardized scoring tool, leading to inconsistent interpretation. AIM: This study aimed to develop and validate a reliable causality assessment tool for AEs associated with medical devices. MATERIALS AND METHODS: A nine-question structured scale was developed from a literature review to classify causality as certain, probable, possible, or unlikely. Validity was established through expert comparisons using the European Union Causality Assessment Scale. Sensitivity, specificity, and area under the curve (AUC) were calculated. Reliability testing occurred in two phases using 50 random Manufacturer and User Facility Device Experience database cases (2014-2024) assessed by six raters (physicians, biomedical engineers, and pharmacists). After 6 weeks, the same cases were reassessed. Reliability was determined using Cronbach's alpha and intraclass correlation coefficient (ICC). RESULTS: The ICC values for phases 1 and 2 were 0.324 and 0.399, respectively (P < 0.001), indicating moderate agreement. Validity testing showed a sensitivity of 90.95%, specificity of 83.25%, and AUC of 0.8605, demonstrating strong accuracy in determining causality. The scale demonstrated high reliability (Cronbach's alpha = 0.893). CONCLUSION: The validated causality assessment tool enhances consistency and reduces subjectivity among evaluators. It provides a reliable scoring system to support regulatory decisions, improve safety, and enable postmarket surveillance. This tool enables systematic assessment of device-related AEs across healthcare settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Challenges and dynamics in reporting medical device incidents: a qualitative studyFrontiers in Health Services · 2025
- Patient safety in dentistry: Surveillance of adverse events associated with medical devices.Journal of dentistry · 2026
- Evaluation of the accuracy, completeness, and compliance of current summary reporting of adverse events to the Food and Drug Administration using the manufacturer and user facility device experience (MAUDE) database: cross sectional study.BMJ (Clinical research ed.) · 2026
- A Nationwide Descriptive Epidemiological and Text-mining Analysis of Incidents and Adverse Events Involving Clinical Engineers in Japan.Journal of patient safety · 2026
- Proof of concept validation of a novel button-enabled electronic adherence monitoring device: An iterative approach combining technical performance and user experience.Research in social & administrative pharmacy : RSAP · 2026
- Standard Elements in Studies of Adverse Events and Medical Error: the SESAME statement.Pediatric quality & safety · 2026