Proof of concept validation of a novel button-enabled electronic adherence monitoring device: An iterative approach combining technical performance and user experience.
Micheline C Sarbach, Muhammed B Karahan, Kirstin Messner, Samuel S Allemann
Pharmaceutical Care Research Group, Department Pharmaceutical Sciences, University of Basel, Klingelbergstrasse 50, 4056, Basel, Switzerland. Electronic address: Micheline.sarbach@unibas.ch. · Pharmaceutical Care Research Group, Department Pharmaceutical Sciences, University of Basel, Klingelbergstrasse 50, 4056, Basel, Switzerland.
Key points
From the paper's conclusion- After iterative optimization, the BEAM device demonstrated high validity, technical reliability and strong usability, supporting its use in clinical research and practice. The findings highlight the importance of combining technical and user-centred evaluation in EM device validation.
Abstract
BACKGROUND AND AIM: Electronic adherence monitoring (EM) devices are increasingly used in clinical research and practice. However, standardized validation approaches remain limited. This study validated a novel button-enabled electronic adherence monitoring device (BEAM) using technical performance testing and user-centred evaluation in an iterative development process. METHODS: A validation protocol was applied in two consecutive test runs. In each run, 20 participants used 20 BEAM devices for 14 days in a simulated medication-taking routine and documented intake events in a paper-based diary. We assessed device performance using sensitivity, specificity, positive predictive and negative predictive values. User experience was evaluated using the System Usability Scale (SUS) and an mHealth satisfaction questionnaire. Standardized tests examined device robustness under mechanical stress and environmental exposure in climate chambers. RESULTS: In the first test run, the BEAM devices showed high specificity (0.99, 95% CI 0.97-0.99) but moderate sensitivity (0.71, 95% CI 0.62-0.79), indicating systematic under-recording. Following technical modifications, performance improved substantially in the second test run, with sensitivity, specificity, positive predictive and negative predictive values all approaching 1.00. Usability remained high throughout both test runs with mean SUS score of 85.8 ± 11.6 and 96.8 ± 7.5. Climate chamber experiments demonstrated stable temperature and relative humidity measurements. Physical tests indicated preserved device functionality after mechanical stress. CONCLUSION: After iterative optimization, the BEAM device demonstrated high validity, technical reliability and strong usability, supporting its use in clinical research and practice. The findings highlight the importance of combining technical and user-centred evaluation in EM device validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Medication adherence in clinical trials: measurement methods, methodological challenges, and reporting standards.Expert review of pharmacoeconomics & outcomes research · 2026
- Development and validation of a standardized causality assessment tool for adverse events associated with medical devices.Perspectives in clinical research · 2026
- SAFER: a simulation-based assessment tool for emergency robotic undocking with multisource validity evidence.Surgical endoscopy · 2026
- Ensemble-based label adjudication (EBLA): A method for validating continuous physiological measures against observer-dependent reference standards.Physiological reports · 2026
- Cross‐Cultural Adaptation and Psychometric Validation of the Arabic User Satisfaction Evaluation Questionnaire (USEQ)Advances in Human-Computer Interaction · 2026
- Feedback-enhanced electronic monitoring for asthma inhaler adherence: A systematic review and meta-analysis of randomized controlled trials.International journal of nursing studies · 2026