Proof of concept validation of a novel button-enabled electronic adherence monitoring device: An iterative approach combining technical performance and user experience.

Micheline C Sarbach, Muhammed B Karahan, Kirstin Messner, Samuel S Allemann

Pharmaceutical Care Research Group, Department Pharmaceutical Sciences, University of Basel, Klingelbergstrasse 50, 4056, Basel, Switzerland. Electronic address: Micheline.sarbach@unibas.ch. · Pharmaceutical Care Research Group, Department Pharmaceutical Sciences, University of Basel, Klingelbergstrasse 50, 4056, Basel, Switzerland.

Key points

From the paper's conclusion
  • After iterative optimization, the BEAM device demonstrated high validity, technical reliability and strong usability, supporting its use in clinical research and practice. The findings highlight the importance of combining technical and user-centred evaluation in EM device validation.

Abstract

BACKGROUND AND AIM: Electronic adherence monitoring (EM) devices are increasingly used in clinical research and practice. However, standardized validation approaches remain limited. This study validated a novel button-enabled electronic adherence monitoring device (BEAM) using technical performance testing and user-centred evaluation in an iterative development process. METHODS: A validation protocol was applied in two consecutive test runs. In each run, 20 participants used 20 BEAM devices for 14 days in a simulated medication-taking routine and documented intake events in a paper-based diary. We assessed device performance using sensitivity, specificity, positive predictive and negative predictive values. User experience was evaluated using the System Usability Scale (SUS) and an mHealth satisfaction questionnaire. Standardized tests examined device robustness under mechanical stress and environmental exposure in climate chambers. RESULTS: In the first test run, the BEAM devices showed high specificity (0.99, 95% CI 0.97-0.99) but moderate sensitivity (0.71, 95% CI 0.62-0.79), indicating systematic under-recording. Following technical modifications, performance improved substantially in the second test run, with sensitivity, specificity, positive predictive and negative predictive values all approaching 1.00. Usability remained high throughout both test runs with mean SUS score of 85.8 ± 11.6 and 96.8 ± 7.5. Climate chamber experiments demonstrated stable temperature and relative humidity measurements. Physical tests indicated preserved device functionality after mechanical stress. CONCLUSION: After iterative optimization, the BEAM device demonstrated high validity, technical reliability and strong usability, supporting its use in clinical research and practice. The findings highlight the importance of combining technical and user-centred evaluation in EM device validation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Proof of concept validation of a novel button-enabled electronic adherence monitoring device: An iterative approach combining technical performance and user experience — Science Skim