Evaluation of the accuracy, completeness, and compliance of current summary reporting of adverse events to the Food and Drug Administration using the manufacturer and user facility device experience (MAUDE) database: cross sectional study.
Alyssa Zhou, Madris Kinard, Sarah Tingley, Rita F Redberg
School of Medicine, University of California, San Francisco, San Francisco, CA, USA. · Device Events, York, PA, USA.
Key points
From the paper's conclusion- Adverse event summary reports in MAUDE are frequently incomplete, inaccurate, and delayed. Improvements are needed to ensure timely identification of device safety signals and uphold FDA transparency standards.
Abstract
OBJECTIVES: To evaluate the accuracy and completeness of adverse event summary reports for medical devices and compliance of the reports with the Food and Drug Administration (FDA) requirements. DESIGN: Cross sectional study. SETTING: The FDA manufacturer and user facility device experience (MAUDE) database. PARTICIPANTS: All adverse event summary reports filed in MAUDE from July 2019 to June 2025. MAIN OUTCOME MEASURES: Identification of device problem codes, appropriate identification as summary reports in the submission form, concordance between the number of events data field and the report narrative, and compliance with FDA reporting deadlines. RESULTS: 6308 summary reports were submitted to MAUDE containing 159 503 adverse events. Use of summary reports increased from 808 to 1757 between 2020 and 2024, a factor of 1.21 per year (P<0.001). Most (91%) adverse events in summary reports were for class 3 high risk devices. Summary reports lacked descriptions of potential device related problems: 57% of reports denoted an "adverse event without identified device or use problem" and 38% of reports selected "insufficient information" in the device problem code data field. Manufacturers incorrectly checked "N" (no) in the summary report (Y/N, reporting yes/no) field for 694 summary reports (11%) and erroneously listed "n=1" under the number of events field for 423 summary reports (7%), causing an undercount of 11 902 adverse events. The majority (55%) of summary reports failed to comply with the FDA reporting deadline of 30 days after the manufacturer became aware of the adverse event. CONCLUSIONS: Adverse event summary reports in MAUDE are frequently incomplete, inaccurate, and delayed. Improvements are needed to ensure timely identification of device safety signals and uphold FDA transparency standards.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Post-market safety signals of the Yomi dental robotic system: an analysis of FDA MAUDE reports.Journal of robotic surgery · 2026
- Adverse event reporting patterns for endodontic files in the FDA MAUDE database: a postmarket surveillance analysis, 2015-2025.Journal of endodontics · 2026
- National adverse event reporting for medical devices in Israel, 2013-2024.Israel journal of health policy research · 2026
- MedDRA adoption and adverse event reporting quality in gastrointestinal and abdominal surgery randomised controlled trials: a cross-sectional analysis.BMJ open · 2026
- Manufacturer and User Facility Device Experience (MAUDE)-Dash: An open-source, interactive dashboard for real-time post-market surveillance of medical devicesSurgical Neurology International · 2026
- Patient safety in dentistry: Surveillance of adverse events associated with medical devices.Journal of dentistry · 2026