Assessment of Calculation Grid Size Impact on Plan Quality and Gamma Evaluation Metrics in Intracranial Stereotactic Radiotherapy.
Ravindra Shende, Joyce Alexis, R S Rahul
Department of Radiation Oncology, Balco Medical Centre, New Raipur, Chhattisgarh, India. · Department of Radiation Oncology, Optimus Oncology Solapur Cancer Centre, Solapur, Maharashtra, India.
Key points
From the paper's conclusion- Study highlights that dose grid size variations significantly influence dosimetric accuracy, plan quality, and QA robustness in intracranial SRT. Coarser calculation grids compromise target coverage and gradient fidelity, while finer grids (≤1.25 mm) preserve near-reference dosimetric accuracy. Adopting a finer calculation grid ensures reliable dose computation, consistent patient-specific QA, and optimal clinical performance in SRT.
Abstract
BACKGROUND: This study aimed to evaluate the impact of dose calculation grid size on dosimetric accuracy, plan quality indices, and patient-specific quality-assurance (QA) outcomes in intracranial stereotactic radiotherapy (SRT). MATERIALS AND METHODS: Fifteen retrospective intracranial SRT plans treated with HyperArc volumetric modulated arc therapy were re-planned using five dose calculation grid sizes of 1.0, 1.25, 1.5, 1.75, and 2.0 mm in the Eclipse treatment planning system. All plans were made to deliver 30 Gy in 5 fractions using 6 MV flattening-filter-free beams and calculated using the Anisotropic analytical algorithm. Plan quality was evaluated using multiple plan quality metrics, including RTOG quality coverage index (QI), Paddick conformity index, homogeneity index, and gradient index (GI). Dose spillage outside the target was assessed using volumetric dose indices. Patient-specific QA was performed using SRS MapCHECK with gamma-analysis under multiple criteria. Statistical comparisons were conducted using dependent paired t-tests, repeated-measure analysis of variance, and Pearson correlation coefficient analysis. RESULTS: Increasing the dose calculation grid size resulted in a systematic increase in planning target volume mean dose and a corresponding reduction in D95%, indicating loss of target coverage attributable to spatial dose averaging. GI showed the strongest dependence on grid size, demonstrating progressive degradation of dose fall-off with coarser grids. Low-dose spillage to the normal brain and brainstem increased significantly with increasing grid size, while distant organs at risk were comparatively less affected. Gamma passing rates declined progressively with increasing grid size, particularly under the most stringent evaluation criteria. CONCLUSION: Study highlights that dose grid size variations significantly influence dosimetric accuracy, plan quality, and QA robustness in intracranial SRT. Coarser calculation grids compromise target coverage and gradient fidelity, while finer grids (≤1.25 mm) preserve near-reference dosimetric accuracy. Adopting a finer calculation grid ensures reliable dose computation, consistent patient-specific QA, and optimal clinical performance in SRT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of grid size and statistical uncertainty parameters on dose calculations using the Monte Carlo algorithm in the Monaco treatment planning system for liver stereotactic body radiotherapy.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Assessment of Clinical Quality Metrics for IROC Houston SRS Phantom Performance EvaluationDigital Commons-TMC (Texas Medical Center) · 2026
- Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy sourceEPJ Web of Conferences · 2026
- ROC-based evaluation of tight gamma criteria for detecting MLC and jaw positioning errors in brain SRS using SRS MapCHECK.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Dosimetric comparison of the Anisotropic Analytical Algorithm and Acuros XB for single-fraction cranial stereotactic radiosurgery with a 6 MV flattening filter-free beam.Biomedical physics & engineering express · 2026
- Assessing Gamma Index Sensitivity to Selected Linear Accelerator Mechanical Errors Using Various Dosimetric Systems.Journal of clinical medicine · 2026