Key points
From the paper's conclusion- All three software platforms provide clinically acceptable independent dose verification for Gamma Knife SRS, with deviations well below AAPM TG-178 recommended tolerance levels. Selection of the appropriate secondary dose calculation platform should be based on institutional workflow integration and specific clinical requirements rather than statistical differences alone.
Abstract
PURPOSE: Independent dose calculation is recommended for stereotactic radiosurgery (SRS) quality assurance. This study evaluates the accuracy of three commercial secondary dose calculation software programs-MUCheck, RadCalc, and ClearCalc-against the Elekta Leksell Gamma Plan (LGP) treatment planning system (TPS) for Gamma Knife SRS. MATERIALS AND METHODS: Dose calculations for 133 patients (239 intracranial targets) planned with the LGP TPS using the TMR10 algorithm were retrospectively analyzed with MUCheck, RadCalc, and ClearCalc. Point dose calculations were compared between each software and LGP. Statistical analysis included analysis of variance (ANOVA) and paired t-tests, with clinical tolerance assessed against the American Association of Physicists in Medicine (AAPM) Task Group (TG)-178 recommendations. RESULTS: Mean differences were 0.34% ± 1.10% (MUCheck), 0.53% ±0.53% (RadCalc), and 0.63% ± 0.16% (ClearCalc). All showed statistically significant differences from LGP but remained within clinical tolerance. Clinical tolerance analysis showed 97.9%, 100%, and 100% of calculations within ±3% for MUCheck, RadCalc, and ClearCalc, respectively. ANOVA revealed significant differences among platforms (F = 10.46, P < 0.001), although all deviations remained clinically acceptable. CONCLUSION: All three software platforms provide clinically acceptable independent dose verification for Gamma Knife SRS, with deviations well below AAPM TG-178 recommended tolerance levels. Selection of the appropriate secondary dose calculation platform should be based on institutional workflow integration and specific clinical requirements rather than statistical differences alone.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- How robust are secondary dose algorithms for Patient-Specific quality assurance in modulated stereotactic Radiosurgery?Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- ROC-based evaluation of tight gamma criteria for detecting MLC and jaw positioning errors in brain SRS using SRS MapCHECK.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Characterization of high-resolution detector response for stereotactic patient-specific QA across different TPS-linac systems.Radiation and environmental biophysics · 2026
- Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery.Radiological physics and technology · 2026
- Beyond Gamma Analysis: An Accessibility-oriented Framework for Comparative Evaluation of Patient-specific Radiotherapy Quality Assurance.Journal of medical physics · 2026
- Assessment of Clinical Quality Metrics for IROC Houston SRS Phantom Performance EvaluationDigital Commons-TMC (Texas Medical Center) · 2026