Commissioning of Medical Linear Accelerators in India: A Survey of Regulatory Compliance, Clinical Implementation, and Professional Challenges.
R C Muhammed Anaz, Shrikant Kale, Ritu Raj Upreti, Raghavendra Hajare, Shubham Gadge, Avdhoot Sutar, Udita Upreti, Yogesh Ghadi,
Department of Medical Physics, Tata Memorial Hospital, Mumbai, Maharashtra, India. · Department of Radiation Oncology, Homi Bhabha Cancer Hospital and Research Centre, Visakhapatnam, Andhra Pradesh, India.
Key points
From the paper's conclusion- Advanced LINAC technologies are being successfully adopted in India; however, greater standardization, structured training, external audit participation, and independent verification frameworks are required to strengthen commissioning quality and patient safety across radiotherapy centers nationwide.
Abstract
PURPOSE: This study evaluated current clinical practices, regulatory compliance, and professional challenges encountered during the commissioning of advanced Linear Accelerators (LINACs) in India. MATERIALS AND METHODS: A nationwide survey using a structured 20-item questionnaire was conducted from August 2025 to March 2026. The questionnaire covered commissioning practices, staffing and vendor support, quality assurance (QA) procedures, clinical implementation, and participant opinions and suggestions. Responses were analyzed to assess trends in technology adoption and variations in commissioning workflows. RESULTS: Data were obtained for 75 LINACs commissioned between 2020 and 2025 across 67 institutions. Frequently commissioned systems included Varian TrueBeam (37.3%), Varian Halcyon (18.7%), and Elekta Versa HD (16%). Advanced treatment techniques such as Volumetric Modulated Arc Therapy (VMAT), Stereotactic Radio Surgery (SRS), and Stereotactic Body Radiotherapy (SBRT) were commonly implemented; however, Total Skin Electron Therapy (TSET) and Total Body Irradiation (TBI) commissioning remained limited. More than one-fourth of the participants (26.7%) had no prior experience in LINAC commissioning. 87.8% participants considered the existing Atomic Energy Regulatory Board's QA requirements adequate, 89.3% supported mandatory inclusion of Treatment Planning System QA within the licensing process. Small-field dosimetry following International Atomic Energy Agency (IAEA) TRS-483 guidelines was performed by 60.8% of centers, while 77.3% had not participated in external dosimetric audits. Major challenges included workforce shortages and balancing commissioning with routine clinical responsibilities. Independent peer review before clinical implementation was recommended by 81.3% of respondents. CONCLUSION: Advanced LINAC technologies are being successfully adopted in India; however, greater standardization, structured training, external audit participation, and independent verification frameworks are required to strengthen commissioning quality and patient safety across radiotherapy centers nationwide.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ACPSEM guideline: commissioning and quality assurance of Magnetic Resonance Imaging Linear Accelerators.Physical and engineering sciences in medicine · 2026
- A pathway away from measurement‐based patient‐specific QA: A review of contemporary quality assurance for external‐beam radiotherapy sub‐systemsJournal of Applied Clinical Medical Physics · 2025
- Six-year data from three magnetic resonance-guided radiotherapy systems demonstrate long-term stability of linear accelerator quality assurance.Physics and imaging in radiation oncology · 2026
- Clinical Experience of Craniospinal Irradiation Using Volumetric Modulated Arc Therapy on the Siddharth-II Linear Accelerator: A Dosimetric Analysis.Cureus · 2026
- Total body irradiation practice patterns: insights from a national survey.Frontiers in oncology · 2026
- Beyond Gamma Analysis: An Accessibility-oriented Framework for Comparative Evaluation of Patient-specific Radiotherapy Quality Assurance.Journal of medical physics · 2026