ACPSEM guideline: commissioning and quality assurance of Magnetic Resonance Imaging Linear Accelerators.
Luis Munoz, Reza Alinaghi-Zadeh, Stephen Gibson, Urszula Jelen, Marcus Powers, Sandra Fisher, Nick Cook, Christopher Colyer,
GenesisCare , Adelaide, SA, Australia. lmunoz@onephysics.com. · Austin Health, Melbourne, VIC, Australia.
Key points
From the paper's conclusion- It was developed by the MRI-Linac QA sub-group of the Magnetic Resonance Imaging Linac Working Group using an evidence-informed, consensus-based approach.
Abstract
MRI-guided radiotherapy introduces unique quality assurance (QA) requirements due to magnetic field effects on dose delivery, imaging-dependent workflows, adaptive planning and system specific considerations. Consistent, vendor agnostic guidance is needed to support safe commissioning, clinical implementation, and ongoing performance monitoring across MRI-Linac platforms. This guideline defines minimum recommended MRI-Linac commissioning and QA practices to support consistent and safe clinical operation across services. It was developed by the MRI-Linac QA sub-group of the Magnetic Resonance Imaging Linac Working Group using an evidence-informed, consensus-based approach. Key international standards, task group reports, vendor documentation, and peer-reviewed literature relevant to the MRI-Linac were reviewed. Candidate QA tests were mapped to clinical risks and system functions, then organized into commissioning and routine QA domains with suggested cadence. Recommendations were refined through iterative working group review, with emphasis on implementation and auditability. The resulting commissioning recommendations are grouped into domains addressing key performance and safety areas for MRI-Linacs. Additionally, a minimum test set is proposed with suggested frequencies and tolerances where supported by evidence. Adoption of this guideline supports harmonized MRI-Linac QA practice, improved reproducibility across sites, and enhanced patient safety. The framework is designed to be implementable across platforms and evolve as MRI-guided radiotherapy workflows, vendor capabilities, and the evidence matures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Six-year data from three magnetic resonance-guided radiotherapy systems demonstrate long-term stability of linear accelerator quality assurance.Physics and imaging in radiation oncology · 2026
- MRI Safety and Disaster Management for MR-Simulator and MR-Linacs.Seminars in radiation oncology · 2026
- Commissioning of Medical Linear Accelerators in India: A Survey of Regulatory Compliance, Clinical Implementation, and Professional Challenges.Journal of medical physics · 2026
- Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgeryJournal of Applied Clinical Medical Physics · 2026
- A novel automatic real-time motion tracking method in MRI-guided radiotherapy using enhanced Tracking-Learning-Detection framework with automatic segmentation.Radiation oncology (London, England) · 2026
- CAR Practice Guidelines on Breast Imaging and Interventions: Breast Magnetic Resonance ImagingCanadian Association of Radiologists Journal · 2026