Reference-based surface agreement of customized implant abutments: a comparison of two workflows.
Jae-Won Kim, Min-Suk Kim, Jung-Bo Huh
Department of Prosthodontics, School of Dentistry, Pusan National University, Yangsan, Republic of Korea. · General Clinic, Pusan National University Dental Hospital, Yangsan, Republic of Korea.
Key points
From the paper's conclusion- Under laboratory conditions, CG showed closer agreement on the measured surfaces, particularly in the molar and premolar conditions and in the axial region. Because the workflows differed in starting stock and machining scope, these findings represent workflow-level differences and do not establish clinical superiority.
Abstract
PURPOSE: This in vitro study compared reference-based surface agreement of customized implant abutments fabricated using an industrial computer numerical control (CNC) machining workflow (CG) and a dental computer-aided design/computer-aided manufacturing (CAD-CAM) milling workflow (MG). MATERIALS AND METHODS: A resin model with molar, premolar, and incisor implant sites was used. Sixty abutments (30 per workflow) were organized into 10 three-abutment scan datasets per workflow. CG specimens were machined from titanium bar stock, including the implant-abutment connection, whereas MG specimens were milled from pre-milled blanks only above pre-machined connections. Scans were aligned to a computer-aided design (CAD)-derived reference using retained non-abutment tooth structures. Deviation was assessed using root mean square (RMS, mm) and in-tolerance rate (inTOL, percentage of points within ± 50 µm). Curve-derived tooth-level RMS was the primary endpoint. RESULTS: Overall surface-based RMS was lower in CG than in MG, whereas inTOL showed near-ceiling values in both groups. In the tooth-level analysis, CG showed lower RMS and higher inTOL in the molar and premolar conditions; neither outcome differed significantly in the incisor condition. In the curve-level analysis, differences favoring CG were concentrated in axial regions, whereas differences favoring MG were limited to molar Curve 2 (RMS and inTOL) and molar Curve 8 (RMS only). CONCLUSION: Under laboratory conditions, CG showed closer agreement on the measured surfaces, particularly in the molar and premolar conditions and in the axial region. Because the workflows differed in starting stock and machining scope, these findings represent workflow-level differences and do not establish clinical superiority.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Retrospective analysis of CAD-CAM occlusal devices.The Journal of prosthetic dentistry · 2026
- Clinical evaluation of the accuracy of static interocclusal single screw-retained implant records designed by a CAD-CAM system: A comparative in vivo study.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Comparative analysis of stability and removal torque of custom CAD-CAM versus stock abutments before and after cyclic loading: A systematic review with meta-analysis and GRADE analysis.The Journal of prosthetic dentistry · 2026
- Does manufacturing specialization matter? Transfer accuracy of static surgical guides produced by general-purpose and dental-specific additive workflows: A paired in vitro study.Journal of clinical and experimental dentistry · 2026
- Influence of impression technique and screw access hole configuration on distortion in conventional impressions of prepared CAD-CAM abutment geometries.The journal of advanced prosthodontics · 2026
- Implant Position Accuracy of a Dual-Function Bite Holder for Single-Visit Registration of Complete-Arch Implants and Maxillomandibular Relationship: An In Vitro Study.The International journal of oral & maxillofacial implants · 2026