The effect of intraoral scanner tip size and scanning strategy on the digital impression trueness of full-arch implant-supported prostheses: an in vitro study.
Aslıhan Öztürk, Ahmet Atila Ertan
Department of Prosthodontics, Faculty of Dentistry, Hacettepe University, Ankara, Turkey.
Key points
From the paper's conclusion- Scanner head size and scanning strategy significantly affected trueness. Clinically, larger scanner heads combined with appropriate scanning strategies may improve trueness in full-arch or extended edentulous scans by reducing cumulative stitching errors.
Abstract
PURPOSE: The purpose of this in vitro study was to evaluate the effect of scanner head size, scanning path, and scanning strategy on the trueness of complete-arch digital implant impressions in an edentulous maxilla. MATERIALS AND METHODS: An edentulous maxillary model with four implants was digitized using a laboratory scanner to generate a reference dataset. Digital impressions were obtained using two scanner head sizes, two scanning paths, and three scanning strategies, resulting in twelve experimental groups, with ten scans for each group under standardized conditions. Root mean square and angular deviations between the reference dataset and each test scan were calculated using a best-fit algorithm. Statistical analysis was performed with the level of significance set at α = .05. RESULTS: The highest median deviation was observed in the medium scanner head with the single scanning strategy group (M.I.S., 51.5 µm), whereas the lowest deviation was found in the large scanner head with the double scanning strategy group (L.P.D., 17.5 µm). All large-head scans showed significantly lower RMS and angular deviation values than the medium-head scans (P < .001). Among medium-head groups, the reverse double scan produced the lowest RMS values (M.I.R., 40.5 µm; M.P.R., 38 µm), while the single-scan groups showed the highest deviations (M.I.S., 51.5 µm; M.P.S., 49.5 µm). Scanning path showed no significant effect on trueness. CONCLUSION: Scanner head size and scanning strategy significantly affected trueness. Clinically, larger scanner heads combined with appropriate scanning strategies may improve trueness in full-arch or extended edentulous scans by reducing cumulative stitching errors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of a stackable scanning guide, intraoral scanner type, and operator on the accuracy of maxillary full-arch implant scans: An in vitro studyJournal of Dentistry · 2026
- Scanner-Dependent Influence of Palatal Surface Acquisition on Trueness, Precision and Transverse Arch Reproduction During Maxillary Complete-Arch Intraoral Scanning.Journal of dentistry · 2026
- Influence of posterior adjacent tooth presence, implant depth, and scanning strategies on trueness in partial edentulism.BMC oral health · 2026
- Influence of margin level and finish line geometry on the trueness of digital scans.The journal of advanced prosthodontics · 2026
- Influence of arch form on the accuracy of digital workflow in all-on-four implant rehabilitation cases (an in vitro study).BMC oral health · 2026
- Accuracy of 12 Intraoral Scanners and Photogrammetric Systems for Full-Arch Edentulous Maxillae: An In Vitro Study.The International journal of periodontics & restorative dentistry · 2026