Key points
From the paper's conclusion- Posterior adjacent tooth presence, implant depth, and scanning strategy significantly influenced trueness. While straight scanning strategies maintained deviations within clinically acceptable limits, deep implants in edentulous terminal areas scanned with zigzag protocols exceeded the 100-150 μm linear threshold, representing the worst-case clinical scenario.
Abstract
BACKGROUND: To evaluate the trueness of digital implant impressions in partially edentulous models according to posterior adjacent tooth presence, implant depth, and intraoral scanning strategy. METHODS: Two maxillary models (Kennedy Class II Mod I) with implants at 1 mm (M1) and 3 mm (M3) subgingival depths were used. The left side retained the second molar, whereas the right side was edentulous. A confocal intraoral scanner was used with 4 strategies: right and left straight (RS, LS), and right and left zigzag (RZ, LZ). Each strategy was repeated 10 times (n = 80). Laboratory scans served as reference. Trueness was evaluated by root mean square (RMS), linear deviation of scan bodies, and inter-implant distance. Data were analyzed using two-way ANOVA (α = 0.05). RESULTS: The presence of a posterior adjacent tooth resulted in lower deviations than its absence (This finding was obtained from the comparison of scanbody (SB) 16 and SB 26; statistically significant differences were observed for RMS in M1 RZ (p < 0.001) and M1 LZ (p = 0.002), and for linear deviation in M1 LZ (p < 0.001), M3 LS (p = 0.005), and M3 LZ (p = 0.005)). Implants placed at 3 mm depth showed significantly higher deviations than those placed at 1 mm depth (p < 0.05). Straight scanning strategies resulted in lower deviations than zigzag strategies (p < 0.01). CONCLUSIONS: Posterior adjacent tooth presence, implant depth, and scanning strategy significantly influenced trueness. While straight scanning strategies maintained deviations within clinically acceptable limits, deep implants in edentulous terminal areas scanned with zigzag protocols exceeded the 100-150 μm linear threshold, representing the worst-case clinical scenario.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The effect of intraoral scanner tip size and scanning strategy on the digital impression trueness of full-arch implant-supported prostheses: an in vitro study.The journal of advanced prosthodontics · 2026
- Digitization of a Single Tissue-Level Implant Without Scan Body - A Feasibility Study.Journal of dentistry · 2026
- Trueness of sequential and simultaneous capture strategies for adjacent implant scan bodies.Journal of dentistry · 2026
- Accuracy of Direct Intraoral Versus Reverse Scanning for Abutment-Level Implant Impressions: an In Vitro Study.European journal of dentistry · 2026
- Evaluation of four digital methods in capturing complete-arch implant positions: Direct scanning, splinted scan bodies, three-dimensional-printed scan aids, and intraoral photogrammetry.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- In vivo deviation of residual ridge between digital and conventional impressions in the partially edentulous maxilla: a pilot study.PeerJ · 2026