Double-Port Totally Endoscopic Mitral Valve Replacement Through a Prosthesis-Sized Working Incision with Posterior Subvalvular Preservation.
Cao D Khang, Tran D Quang, Phan Q Thuan, Vuong N Minh, Pham T V Chuong, Nguyen H Dinh
Department of Cardiovascular Surgery, University Medical Center, University of Medicine and Pharmacy at Ho Chi Minh City, Ho Chi Minh City, Vietnam. · Department of Cardiovascular and Thoracic Surgery, School of Medicine, University of Medicine and Pharmacy at Ho Chi Minh City, Ho Chi Minh City, Vietnam.
Key points
From the paper's conclusion- The double-port DCT approach was technically feasible and was associated with acceptable early clinical and echocardiographic outcomes. Because of the retrospective single-center design, small sample size, absence of a comparator group, and short follow-up, these findings should be interpreted as feasibility data rather than evidence of superiority.
Abstract
OBJECTIVE: To evaluate the feasibility and early outcomes of totally endoscopic non-robotic mitral valve replacement using a double-port DCT approach with posterior subvalvular preservation. METHODS: In this single-center retrospective observational study, 76 consecutive patients underwent totally endoscopic non-robotic mitral valve replacement between November 2023 and December 2025. The DCT approach was defined by exclusive endoscopic visualization, no rib spreading, a camera port and prosthesis-sized working incision placed in the same intercostal space, and thoracoscopic intracardiac instrumentation. Posterior ventricular-annular continuity was preserved whenever feasible by posterior leaflet preservation, native chordal preservation, or artificial chordal reimplantation after leaflet resection. Outcomes were reported descriptively with 95% confidence intervals. RESULTS: Mechanical prostheses were implanted in 46 patients (60.5%) and bioprostheses in 30 (39.5%). Mean skin incision length was 3.8 ± 0.5 cm. Mean cardiopulmonary bypass and aortic cross-clamp times were 148.3 ± 38.6 and 101.9 ± 32.0 minutes, respectively. There was 1 early non-cardiac death (1.3%; 95% CI, 0.03%-7.1%), no late death, and no mitral valve reoperation (0%; 95% CI, 0%-4.7%). No moderate or severe paravalvular regurgitation was observed during early echocardiographic follow-up. Mean follow-up was 9.7 ± 6.0 months. Freedom from prosthetic valve-related adverse events was 98.6% (95% CI, 90.6%-99.8%) at 6 months and 95.2% (95% CI, 80.1%-98.9%) at 12 months. The 24-month estimate was identical but was based on few patients at risk and should be interpreted cautiously. CONCLUSIONS: The double-port DCT approach was technically feasible and was associated with acceptable early clinical and echocardiographic outcomes. Because of the retrospective single-center design, small sample size, absence of a comparator group, and short follow-up, these findings should be interpreted as feasibility data rather than evidence of superiority.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Standardized totally endoscopic non-robotic mitral valve repair for degenerative mitral regurgitation: early single-center outcomes in 84 patients.Global cardiology science & practice · 2026
- Totally endoscopic mitral valve replacement: reasons for non-feasibility of repair and 5-year outcomes.Frontiers in cardiovascular medicine · 2026
- Early Outcomes of 2-Port Totally Endoscopic Mitral Valve Repair with Artificial Chordae for Degenerative Mitral Regurgitation: A Single-Center Experience.Journal of chest surgery · 2026
- Single-Port Robotic Mitral Valve Repair: First-in-Human Technique and 30-Day Outcome.JACC. Case reports · 2026
- Comparative outcomes of totally endoscopic versus direct vision mini-thoracotomy mitral valve repair: A propensity score-matched retrospective analysis.JTCVS open · 2026
- Valve-in-valve transcatheter mitral valve replacement via totally endoscopic endoclamp-assisted approach after failed bioprosthetic mitral valve: a case report.Journal of cardiothoracic surgery · 2026