Analytical Validation of a Multielement ICP-MS Method for Human Serum and Its Application to Trace Element Biomonitoring in Hemodialysis Patients.
Sara Valderrama Sanz, Raquel Sánchez Romero, Ricardo Molina Gasset, José Luis Todolí Torró
Clinical Analysis Service, Hospital Virgen de los Lirios, Polígono de Caramanchel s/n, 03804 Alcoi, Spain. · Analytical Chemistry Department, Nutrition and Food Sciences, University of Alicante, Carretera San Vicente del Raspeig s/n, 03690 San Vicente del Raspeig, Spain.
Key points
From the paper's conclusion- This study describes the development and analytical validation of a multielement inductively coupled plasma mass spectrometry (ICP–MS) method for the simultaneous determination of essential and potentially toxic trace elements in human serum (Ti, V, Cr, Mn, Co, Ni, Cu, Zn, As, Cd, Ba, Bi) using a simple direct-dilution procedure with 1% (v/v) nitric acid.
Abstract
Reliable determination of trace elements in human serum is essential for clinical assessment, toxicological investigations and human biomonitoring. This study describes the development and analytical validation of a multielement inductively coupled plasma mass spectrometry (ICP-MS) method for the simultaneous determination of essential and potentially toxic trace elements in human serum (Ti, V, Cr, Mn, Co, Ni, Cu, Zn, As, Cd, Ba, Bi) using a simple direct-dilution procedure with 1% (v/v) nitric acid. Method validation included assessment of linearity, limits of detection and quantification, carry-over, precision, trueness, measurement uncertainty and analyte stability. Calibration curves showed excellent linearity, with coefficients of determination (R2) ranging from 0.9993 to 1.0000 across analyte-specific calibration ranges of 0-10, 0-50, and 0-500 µg L-1. Repeatability ranged from 0.69% (Cu) to 14.5% (Ba), while intermediate precision ranged from 1.32% (Zn) to 81.1% (Cd and Bi at low concentrations). Recoveries obtained using certified serum reference materials ranged from 55% (V) to 206% (Bi) across the evaluated conditions. Meanwhile, limits of detection ranged from 0.01 (V) to 0.16 (Cu) µg L-1. Measurement uncertainty was estimated following the EURACHEM/CITAC and GUM recommendations, identifying repeatability as the principal contributor to the overall uncertainty budget. The method was successfully applied to serum samples from healthy individuals and patients undergoing chronic hemodialysis. The proposed methodology combines simple sample preparation with comprehensive metrological characterisation, providing practical approach multielement serum analysis and its application to trace-element biomonitoring investigations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A systematically evaluated SEC-ICP-MS/MS workflow for Mn, Fe, Cu, and Zn speciation in human serum.Talanta · 2026
- B-052 Automation of sample preparation workflow for trace elements testing by ICP-MS/MSClinical Chemistry · 2026
- Ion chromatography for inorganic phosphate in human serum: a candidate reference measurement procedure and commutable certified reference material.Analytical and bioanalytical chemistry · 2026
- Whole-Blood Trace Elements in Archived Head and Neck Squamous Cell Carcinoma and Source-Labelled Comparison Samples: An Exploratory Analysis with Analytical-Quality Audits.Biological trace element research · 2026
- A-194 Development of candidate Standard Reference Material® 1950a Metabolites in Frozen Human PlasmaClinical Chemistry · 2026
- B-416 Validation of Trace Elements Concentrations in Commonly Used Urine ContainersClinical Chemistry · 2026