Clinical Chemistry

B-416 Validation of Trace Elements Concentrations in Commonly Used Urine Containers

MaryAnn Udechukwu, Jessica Boyd

Key points

From the paper's conclusion
  • However, elements such as selenium, zinc, and lead in some urine jugs may significantly contribute to measured concentrations, emphasizing the need for specimen collection in trace element-free containers.

Abstract

Background Laboratory measurement of trace elements in urine is essential for identifying toxic element exposures and monitoring treatment. Preanalytical contamination from specimen collection containers can lead to falsely elevated results and inappropriate clinical interpretations and decisions. Although specimen collection in trace element-free containers is recommended, limited data exist on the elemental contribution of commonly used commercial urine cups and jugs. This study evaluated trace element concentrations in widely used urine collection containers. Methods One or more lots of urine cups and 24-hour urine jugs were obtained from multiple commercial manufacturers. Urine cups from Greiner Bio-One, BD Vacutainer®, Fisherbrand™, VWR®, and Sarstedt, and 24-hour urine jugs from Fisherbrand™, and Sarstedt were evaluated. The containers were filled with either 2% nitric acid or clinical laboratory reagent water and incubated for 24 hours at room temperature to extract soluble elements. Concentrations of arsenic, cadmium, cobalt, chromium, copper, mercury, iodine, manganese, nickel, lead, selenium, thallium, and zinc were measured using inductively coupled plasma-mass spectrometry. Measured concentrations were compared with the lower limit of quantitation (LLOQ) for each analyte. Results Across all containers evaluated, most trace elements were either undetectable or present at very low concentrations. Urine cups demonstrated minimal elemental contribution compared with 24-hour urine jugs, with most analytes showing concentrations below or equal to 0.07 µg/dL. Urine jugs exhibited greater concentrations for select elements, including chromium (less than or equal to 0.20 µg/dL), iodine (less than or equal to 0.20 µg/dL), arsenic (less than or equal to 1.00 µg/dL), lead (less than or equal to 1.70 µg/dL), zinc (0.30-1.70 µg/dL), and selenium (0.60-3.60 µg/dL). Selenium was consistently detected in both urine cups and jugs and demonstrated the highest concentrations among all elements tested. All measured element concentrations remained below the LLOQ for the analytical method. Lastly, trace elements concentrations did not differ between urine container lots within manufacturers. Conclusion Commonly used commercial urine cups and jugs demonstrated low trace element contributions, supporting their general suitability for urine trace element testing. However, elements such as selenium, zinc, and lead in some urine jugs may significantly contribute to measured concentrations, emphasizing the need for specimen collection in trace element-free containers. Periodic verification of different urine container lots is crucial to ensure trace elements concentrations remain within acceptable limits to avoid specimen contamination.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers