Agreement of 11-Sensor Plantar-Pressure Configurations with a Full Pedar-X Array for Estimating Center of Pressure and Pressure-Derived Vertical Force in Individuals with Flatfoot During Walking and Running.

Min-Yan He, Ping-Hung Lee, Tzyy-Yuang Shiang, Yi-Jia Lin

School of Physical Education, Jiangsu University of Technology, Changzhou 213000, China. · Department of Athletics Performance, National Taiwan Normal University, Taipei 10610, Taiwan.

Key points

From the paper's conclusion
  • This study evaluated the agreement between the center of pressure (CoP) and pressure-derived vertical ground reaction force (vGRF) estimated from minimal 11-sensor configurations and a full Pedar-X array in individuals with flatfoot during walking and running.

Abstract

Dense in-shoe pressure arrays (e.g., 99 sensors) are costly and cumbersome, limiting gait monitoring during dynamic activities. This study evaluated the agreement between the center of pressure (CoP) and pressure-derived vertical ground reaction force (vGRF) estimated from minimal 11-sensor configurations and a full Pedar-X array in individuals with flatfoot during walking and running. Twelve adults with flatfoot performed walking (5 and 7 km/h) and running (11 and 13 km/h) trials with simultaneous Pedar-X recording. Four 11-sensor configurations were compared against the 99-sensor full-array reference using Pearson correlation, root mean square error (RMSE), two-way random-effects intraclass correlation coefficients ICC(A,1), and Bland-Altman analysis. All configurations correlated significantly with the full-array reference for CoP (r = 0.859-0.971) and vGRF (r = 0.867-0.990). CoP RMSE ranged from 5.23 to 8.13 mm (medial-lateral) and 14.00 to 28.20 mm (anterior-posterior), with ICCs of 0.66-0.74 and 0.84-0.92, respectively. Raw vGRF showed moderate agreement (ICC = 0.46-0.56) with systematic underestimation, whereas per-user proportional calibration improved agreement to ICC = 0.90-0.98. Carefully placed minimal sensor configurations can track CoP and the vGRF waveform in individuals with flatfoot with high agreement relative to a full-array reference, supporting low-cost wearable gait monitoring with brief user calibration.

Read the paper