Sensors (Basel, Switzerland)

Multilayer CaF2 Prism-Coupled Ag/Si3N4/Al2O3 SPR Biosensor for the Early Detection of Carcinoembryonic Antigen Biomarker: A Numerical Approach.

Mujeeb Ur Rehman Laghari, Abdul-Aleem Jamali, Xin-Cheng Ren, Fatima Qureshi, Masood Ali Koondher, Muhammad Rizwan Anjum, Atta Muhammad Panhyar

Department of Electronic Engineering, Quaid-e-Awam University of Engineering, Science and Technology, Nawabshah 67480, Sindh, Pakistan. · Department of Biomedical Engineering, Quaid-e-Awam University of Engineering, Science and Technology, Nawabshah 67480, Sindh, Pakistan.

Key points

From the paper's conclusion
  • The results demonstrated that the optimized sensor structure achieved stability, with the highest sensitivity reaching up to 361.08°/RIU for different concentrations and the FoM reaching 145.85 RIU−1.

Abstract

The diagnostic tool for the detection of cancer at the initial stage remains the challenging task for the biosensing community, as it requires the detection of a lower concentration (such as <5 ng/mL) of cancer biomarkers, i.e., the carcinoembryonic antigen. For this purpose, the optimization of the multilayered surface plasmon resonance (SPR) nano-sensor configuration is presented in this paper. The optimized sensor comprises the CaF2 prism, silver (Ag), Si3N4, and Al2O3 nanolayers configured in a Kretschmann setup. An electromagnetic analysis is performed to evaluate the performance parameters of the proposed biosensor in CST Studio using the finite integration technique, and the results are also validated through the transfer matrix method. This includes the computation of sensitivity and the figure of merit (FoM). Systematic optimization was performed for the thickness of functional layers. The results demonstrated that the optimized sensor structure achieved stability, with the highest sensitivity reaching up to 361.08°/RIU for different concentrations and the FoM reaching 145.85 RIU-1. It was significant to observe that the optimized configuration remains stable across the full concentration range, while the configuration with thick Si3N4 undergoes a resonance degradation mode at high analyte indices. There is substantial potential of the proposed sensor for biomedical applications such as early-stage detection of lung, colon, liver, and breast cancers.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers