A Detuning-Resilient Wideband Microstrip Antenna for Wearable Biomedical Telemetry Through Progressive Interdigital Slot Integration and Patch Defect Engineering.
Prince O Siaw, Ebenezer Adjei, Ahmad Aldelemy, John G Buckley, Raed Abd-Alhameed
Institute of Digital and Sustainable Futures, University of Bradford, Bradford BD7 1DP, UK. · Management, Sciences and Engineering, University of Bradford, Bradford BD7 1DP, UK.
Key points
From the paper's conclusion- The compact size and wide fractional bandwidth provide enhanced resilience against environmental detuning, making this design particularly suitable for robust wearable biomedical telemetry.
Abstract
This paper presents the systematic design, equivalent circuit analysis, and experimental validation of a compact wideband circular microstrip patch antenna for use in biomedical telemetry. The antenna, loaded with interdigital capacitive slots and a defected patch structure (DPS), was fabricated on a Rogers RO3006 substrate (relative permittivity 6.5, loss tangent 0.002, thickness 0.8 mm) with overall dimensions of 25 × 36.96 mm2. A stepwise design evolution through five variants (ANT1-ANT5) illustrates the progressive bandwidth enhancement mechanisms. The final design (ANT5) achieved a simulated -10 dB impedance bandwidth of 1.50 GHz (2.51-4.01 GHz), corresponding to a fractional bandwidth of 46%. The fabricated prototype exhibited a measured -10 dB bandwidth of 1.42 GHz (2.54-3.96 GHz), representing a fractional bandwidth of 43.7%. An equivalent lumped-element circuit model was derived and validated against the full-wave simulation results. The antenna exhibited a simulated peak free-space gain of 3.2 dBi at 3.8 GHz, with free-space radiation efficiency exceeding 90%. Under multi-layer tissue phantom loading, the antenna maintained a robust 910 MHz operational bandwidth with a safe maximum specific absorption rate (SAR) of 0.724 W/kg and an on-body efficiency of 8.4%. The compact size and wide fractional bandwidth provide enhanced resilience against environmental detuning, making this design particularly suitable for robust wearable biomedical telemetry.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Compact dual-band implantable antenna for biomedical telemetry with T – shaped and circular patch resonators featuring capacitive ground loadingResults in Engineering · 2026
- Compact Size CP Broadband Implantable Antenna Using Rectangular Slots.TheScientificWorldJournal · 2026
- A Compact Circularly Polarized Antenna for Wireless Medical ApplicationsMicrowave and Optical Technology Letters · 2026
- A highly isolated compact dual-port implantable antenna for biomedical implant applicationsPhysica Scripta · 2026
- Four-Element Miniaturized Implantable MIMO Antenna Design for Scalp BiotelemetryTechnologies · 2026
- Precision Sensing at Selected Spectrum: A Double-Negative Low-Cost Metamaterial Sensor With Enhanced Directionality for Biomedical ApplicationsIEEE Sensors Journal · 2026