Accuracy, Reliability and Waveform-Level Agreement Between the Impera Force Platform and a Criterion-Standard Force Plate.
Dario Pompa, Alessandra Caporale, Antonio Buglione, Pietro Picerno, Johnny Padulo
Department of Medicine and Aging Sciences, University "G. d'Annunzio" of Chieti-Pescara, 66100 Chieti, Italy. · Department of Neuroscience, Imaging and Clinical Sciences, University "G. d'Annunzio" of Chieti-Pescara, 66100 Chieti, Italy.
Key points
From the paper's conclusion- This study evaluated the accuracy and between-session reliability of a portable force platform (Impera, Spinitalia) under static loading, and its concurrent validity during vertical jumping, against a criterion-standard piezoelectric plate with calibration traceable to national standards (4Jump, Kistler).
Abstract
Laboratory-grade force platforms are the reference standard for quantifying ground reaction forces but remain costly and confined to well-resourced laboratories, limiting field-based use. This study evaluated the accuracy and between-session reliability of a portable force platform (Impera, Spinitalia) under static loading, and its concurrent validity during vertical jumping, against a criterion-standard piezoelectric plate with calibration traceable to national standards (4Jump, Kistler). Accuracy and between-session reliability were assessed against five tared loads (21.6-103.2 kg) applied at five plate locations across two sessions 24 h apart. Concurrent validity was assessed during squat and countermovement jumps performed by ten physically active men, using intraclass correlation coefficients, Bland-Altman analysis and the linear fit method applied to the entire force-time waveform. Both platforms underestimated the lightest load by about 4% and approached zero error at higher loads; the between-device difference was 0.19 kg and did not translate into a difference in proportional accuracy (p = 0.151), with no device × position interaction. Between-session coefficients of variation were below 1% for both devices. Concurrent agreement was good to excellent for all jump variables (ICC 0.867-0.996), and waveform agreement was close to unity (R2 ≥ 0.997), with about 2% amplitude underestimation, greater in participants generating the highest impact forces. The Impera platform provides accurate and reliable measurements under static loading, and force-time data that agree closely with the 4Jump during vertical jump testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Concurrent validity of a dual three-dimensional portable force-plate system against a laboratory force platform across vertical, anteroposterior and mediolateral loading tasks.Frontiers in bioengineering and biotechnology · 2026
- Single joint assessment of isometric plantarflexor force characteristics using portable force plates: A systematic review of methods and reliability.Journal of sports sciences · 2026
- A novel force scaling method for ground reaction force estimation from pressure insoles to enable long-term remote monitoring.PloS one · 2026
- Advancements in Smart Wearable 3D printed Insoles for Ground Reaction Force Estimation and Gait AnalysisJournal of Physics: Conference Series · 2026
- In-Field IMU-Based Monitoring of Running-Specific Prostheses and Athlete Movement During Long-Distance Running: A Case StudyApplied Sciences · 2026
- From pressure to force: a device-independent, predictive model of plantar loading for personalized orthoses.Frontiers in bioengineering and biotechnology · 2026