Initial Validation of Markerless Gait Analysis Using SPLYZA Motion: Waveform Agreement Under Practical Measurement Conditions.
Yu Inoue, Ryo Yamasaki, Koji Ono, Shunsuke Yamashina, Kiyofumi Ando, Hiroyuki Doi
School of Human Science, Kibi International University, Takahashi 716-8508, Japan. · Department of Rehabilitation, Kurashiki Heisei Hospital, Kurashiki 710-0826, Japan.
Key points
From the paper's conclusion- These findings suggest that SPLYZA Motion may serve as a complementary tool for repeated waveform-based gait assessments under practical measurement conditions.
Abstract
This study aimed to conduct an initial validation of markerless gait analysis using SPLYZA Motion by evaluating waveform agreement under practical measurement conditions. Joint angle waveforms were compared with those obtained from a marker-based motion capture system. SPLYZA Motion was found to be in good waveform agreement with the reference system for trunk and proximal lower-limb kinematics. Larger errors were observed in distal joints, particularly under occlusion and higher cadence; however, these differences were phase-dependent and did not substantially affect overall waveform patterns. These findings suggest that SPLYZA Motion may serve as a complementary tool for repeated waveform-based gait assessments under practical measurement conditions. Further validation in larger and more diverse populations is required before broader clinical application can be considered.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of View-Specific Markerless Motion Analysis for Lower-Extremity Kinematics After Anterior Cruciate Ligament ReconstructionSensors · 2026
- Lighting and clothing effects on Theia3D markerless joint-angle waveforms during walking.Journal of biomechanics · 2026
- Markerless motion capture for running: validity and reliability of whole-body, joint, and muscle kinematicsJournal of Biomechanics · 2025
- A Preliminary Concurrent Validation of a Portable Markerless Motion Capture System Against Three-Dimensional Gait Analysis Following Total Hip and Knee Arthroplasty.Sensors (Basel, Switzerland) · 2026
- From linear to nonlinear gait measures: validity of a multi-view markerless motion capture systembioRxiv · 2026
- Evaluating the Agreement of Markerless Motion Capture for Joint Angle Estimation in a Constrained Hallway Setting Compared With a Traditional Laboratory SettingJournal of Applied Biomechanics · 2026