Experimental Determination of Detector-Relevant Properties of ZnSe(Al,O) for Beta Spectroscopy.
Arjana Kolnikaj, Kelum A A Gamage, Olaoluwa Popoola, James Graham, Antonio Di Buono
James Watt School of Engineering, University of Glasgow, Glasgow G12 8QQ, UK. · United Kingdom National Nuclear Laboratory, Sellafield, Seascale, Cumbria CA20 1PG, UK.
Key points
From the paper's conclusion- These results provide an experimentally determined parameter set for this ZnSe(Al,O) crystal and support its use in compact beta spectroscopy, while also highlighting the need for direct material characterisation before reliable detector simulation.
Abstract
ZnSe-based scintillators are promising materials for compact beta spectroscopy, but the detector-relevant properties of ZnSe(Al,O) remain poorly documented and strongly dependent on dopant and defect structure. This work presents an experimental characterisation of a commercially supplied ZnSe(Al,O) crystal intended for an SiPM-based beta detector to Sr-90/Y-90 detection. A scanning electron microscope with energy dispersive x-ray spectroscopy (SEM-EDS) was used to verify the elemental composition and examine surface artefacts, while optical and scintillation properties were determined through density measurement, Brewster-angle refractive index measurement, UV-excited emission spectroscopy, Sr-90/Y-90 charge-spectrum analysis, and waveform decay fitting. The crystal was found to contain Zn, Se, Al, and O, with average atomic abundances of 50.3%, 44.0%, 1.2%, and 4.5%, respectively. The measured density was 5.31 g cm-3 and the refractive index was 2.574 at 650 nm. The emission spectrum peaked at approximately 581 nm, within the visible response range of silicon photodetectors. Sr-90/Y-90 beta measurements gave an estimated intrinsic light yield of approximately 7.69×104 photons MeV-1, while exponential fitting of digitised waveforms gave a decay time of 9.41 μs. These results provide an experimentally determined parameter set for this ZnSe(Al,O) crystal and support its use in compact beta spectroscopy, while also highlighting the need for direct material characterisation before reliable detector simulation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Physics-based modelling of alpha detection in ZnS(Ag) scintillation cells for radon measurements.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Fundamental characterisation of CdZnTe (CZT) and CdTe as sensor materials for the JUNGFRAU charge-integrating ASICJournal of Instrumentation · 2026
- Laser-excited photoluminescence of Cs-137 gamma-preconditioned ZnO on ex vivo chicken skin: Concentration-dependent optical enhancement and Biophotonic Implications.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Temperature Dependence of Fe 2+ Spectroscopic Properties in Chalcogenide Single CrystalsEPJ Web of Conferences · 2026
- 6LiInP2Se6 Single Crystals: Growth, Defect Structure, α-Particle Spectroscopy, and Direct-Conversion X-ray Response.ACS applied materials & interfaces · 2026
- Background characterization in the PIKACHU Phase 1 search for two-neutrino double-beta decay of $^{160}$GdarXiv · 2026