Fundamental characterisation of CdZnTe (CZT) and CdTe as sensor materials for the JUNGFRAU charge-integrating ASIC
Jonathan Mulvey, Dominic Greiffenberg, Bernd Schmitt, Anna Bergamaschi, Martin Brückner, Maria del Mar Carulla, Tomáš Celko, Roberto Dinapoli,
Key points
From the paper's conclusion- Two candidate high-Z sensor materials, cadmium zinc telluride (CZT) and cadmium telluride (CdTe), have undergone fundamental characterisation at PSI using the JUNGFRAU readout ASIC.
Abstract
The upgraded Swiss Light Source (SLS 2.0) increases the need for high-Z detector systems with improved detection efficiency for hard X-rays compared with silicon. Two candidate high-Z sensor materials, cadmium zinc telluride (CZT) and cadmium telluride (CdTe), have undergone fundamental characterisation at PSI using the JUNGFRAU readout ASIC. Two variants of each material were investigated: Schottky and Ohmic CdTe, and spectroscopic and high-flux CZT. Measurements include current-voltage characteristics, pixel-resolved leakage current and noise, flat-field response, and energy resolution. Both CZT sensors exhibited a very low mean leakage current of 0.60 pA and 0.18 pA per pixel for spectroscopic and high-flux CZT, respectively. Ohmic CdTe showed a substantially higher mean leakage current of 65.81 pA per pixel, while Schottky CdTe showed much lower mean leakage current (0.62 pA) but contained a significant population of high-leakage-current pixels. At 59.54 keV, both CZT sensors achieved an energy resolution of approximately 3.5 keV FWHM using 3×3 clusters, compared with 4.3 keV for Ohmic CdTe and 5.9 keV for Schottky CdTe. Ohmic CdTe additionally required approximately two hours after biasing for its leakage current and pedestal to stabilise. Both CZT sensors could be operated at full duty cycle with a 500 μs exposure time, with only a small reduction in available dynamic range. Overall, the CZT sensors showed the most favourable performance of the devices investigated, although further measurements across a larger number of sensors are required to assess sensor-to-sensor variation and production yield.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- CadTool: An End-to-End Semiconductor Detector Simulation ToolCondensed Matter · 2026
- Reduced dark current in Cd0.9Zn0.1Te detector arrays by aluminum oxide passivationApplied Physics Letters · 2026
- Setting-up a new Timepix4-based computed tomography system with a CdTe sensor for cultural heritage applications: challenges and preliminary resultsJournal of Instrumentation · 2026
- Defect-assisted space-charge accumulation drives electric-field reconstruction in cadmium zinc telluride x-ray detectorsApplied Physics Letters · 2026
- Heterovalent Cd2+-Doped Cs3Cu2I5 Scintillators with Modulated Exciton-Phonon Coupling for High-Resolution X-ray Imaging.ACS applied materials & interfaces · 2026
- Charge transport and gain normalization condition the absorber-thickness benefit in a compact Si-CZT detectorJournal of Instrumentation · 2026