Experimental Validation of a Compact and Versatile Bioimpedance Measurement Platform Based on the SENSIPLUS Chip.

Lorenzo Giannini, Rita Asquini, Alessio Buzzin, Simone Contardi, Paolo Bruschi, Emanuele Piuzzi

Department of Information Engineering, Electronics and Telecommunications, Sapienza University of Rome, 00184 Rome, Italy. · Interdisciplinary Department of Wellbeing, Health and Environmental Sustainability, Sapienza University of Rome, 02100 Rieti, Italy.

Key points

From the paper's conclusion
  • This work presents a comprehensive experimental validation of a compact bioimpedance measurement platform based on the SENSIPLUS chip, a CMOS sensor interface integrating a frequency-programmable lock-in amplifier for Electrochemical Impedance Spectroscopy in the 10 kHz-1 MHz range.

Abstract

The growing demand for wearable and Internet of Medical Things (IoMT) devices is driving the development of compact, low-power platforms for continuous physiological monitoring. Bioimpedance analysis represents a versatile non-invasive technique for the assessment of tissue properties, body composition, and respiratory dynamics. This work presents a comprehensive experimental validation of a compact bioimpedance measurement platform based on the SENSIPLUS chip, a CMOS sensor interface integrating a frequency-programmable lock-in amplifier for Electrochemical Impedance Spectroscopy in the 10 kHz-1 MHz range. The platform was validated at three complementary levels: (i) electrical characterization on Debye tissue-equivalent circuits using a three-point bilinear calibration, with analysis of the electrode-skin contribution and repeatability assessment; (ii) in vivo multi-frequency bioimpedance spectroscopy (BIS) with Cole-Cole model fitting and hook-effect correction; and (iii) single-frequency thoracic impedance plethysmography for respiratory monitoring. Results were compared against an Agilent E4980A precision Inductance (L), Capacitance (C), and Resistance (R) meter and a calibrated spirometer. The presented device achieved a maximum resistance error below 5.7% and reactance deviation under 6 Ω across the investigated frequency range, Cole-Cole parameters consistent with reference values, and strong linear correlation (R2=0.97) between thoracic impedance variations and tidal volume, with respiratory rate estimation errors below 2% across the ten sessions, specifically 1.43% during normal breathing and 1.96% during deep breathing. These results demonstrate that the SENSIPLUS-based platform achieves metrological performance compatible with the requirements of wearable IoMT applications, here demonstrated in a single-subject proof-of-concept study, while relying for all critical analog functions on a compact (1.5×1.5) mm2 system-on-chip with low power consumption (1.5 mW).

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers