Accuracy of Dynamic Computer-Assisted Surgery for Pterygoid Implant Placement in Fully and Partially Edentulous Maxillae: A Retrospective Comparative Study.
Luka Plivelić, Ivica Dubravica, Ivan Zajc, Vlatka Debeljak, Ana Zulijani
Faculty of Dental Medicine and Health of Osijek, Josip Juraj Strossmayer University of Osijek, 31000 Osijek, Croatia. · Polyclinic Dubravica, 22211 Vodice, Croatia.
Key points
From the paper's conclusion- This study aimed to compare the accuracy of a dynamic computer-assisted implant surgery system (dCAIS) for pterygoid implant placement in fully and partially edentulous maxillae using radiographic marker registration (RMR) and markerless tracing registration (MTR), respectively.
Abstract
Background and Objectives: The rehabilitation of the atrophic posterior maxilla with pterygoid implant presents a significant challenge due to complex regional anatomy and limited visual access. This study aimed to compare the accuracy of a dynamic computer-assisted implant surgery system (dCAIS) for pterygoid implant placement in fully and partially edentulous maxillae using radiographic marker registration (RMR) and markerless tracing registration (MTR), respectively. Materials and Methods: Forty pterygoid implants were retrospectively evaluated in 40 patients, divided into two groups: fully edentulous (n = 20) and partially edentulous (n = 20). Implant placement was performed using a dynamic navigation system (dCAIS), utilizing either bone-anchored mini-screws or tooth-surface tracing as reference points. Accuracy was assessed by superimposing preoperative plans with postoperative cone beam computer tomography (CBCT) scans. Coronal, apical, and angular deviations were measured and analyzed using the Data Science Workbench (version 14). The level of statistical significance was set at α = 0.05 (two-tailed). Results: The mean deviations for fully and partially edentulous maxillae measured 1.24 ± 0.27 mm and 1.40 ± 0.27 mm at the coronal level (p = 0.065), 1.30 ± 0.49 mm and 1.30 ± 0.54 at the apical level (p = 0.995), and 0.72 ± 0.38° and 0.82 ± 0.43° in angular deviation (p = 0.560), respectively. No statistically significant differences were found between the two groups. Notably, mean angular deviation was remarkably low (<1°) in both groups. Conclusions: No statistically significant between-group differences in accuracy were detected in coronal, apical, and angular deviations between the markerless tracing registration method in partially edentulous patients and the bone-anchored radiographic marker registration method in fully edentulous patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of dynamic computer-assisted implant surgery in fully edentulous patients in a private practice: A retrospective cohort study.Journal of dentistry · 2026
- Comparative accuracy of dynamic navigation versus static surgical guides for dental implant placement: an in vitro assessment of free-end and non-free-end configurationsScientific Reports · 2026
- Accuracy of Pterygoid Implant Placement With Augmented and Mixed Reality Navigation: An In Vitro Study.Clinical oral implants research · 2026
- The Pterygoid Anchorage Spatial Adequacy Concept: A CT-Based Morphometric Framework for Pterygoid Implant Planning.Journal of clinical medicine · 2026
- Effect of Alveolar Ridge Inclination on Dynamic Navigation Accuracy of Implant Surgery: A Retrospective Clinical Study.Clinical, cosmetic and investigational dentistry · 2026
- Prospective analysis of site-specific accuracy and error sources in dynamically navigated implant surgery.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026