Patient-Clustered Analysis of Recorded Insertion Torque in 101 MultiNeO Titanium Dental Implant Placements: An Exploratory Retrospective Cohort Study.
Katarzyna Wieczorek, Adrian Wasilewski, Bartłomiej Szwed, Weronika Nicpoń, Natalia Szymkowiak, Grzegorz Hajduk, Mansur Rahnama-Hezavah, Michał Łobacz
Chair and Department of Oral Surgery, Medical University of Lublin, ul. Chodźki 6, 20-093 Lublin, Poland.
Key points
From the paper's conclusion- This retrospective single-centre study included 101 MultiNeO titanium implant placements in 55 adults treated between April 2023 and January 2026.
Abstract
Primary implant stability arises from mechanical engagement at placement, and insertion torque is not a direct measure of osseointegration. This retrospective single-centre study included 101 MultiNeO titanium implant placements in 55 adults treated between April 2023 and January 2026. Four experienced operators placed implants subcrestally using a W&H Implantmed SI-1023, a bone-resistance-adapted drilling protocol, and a programmed target of 35 N·cm. Associations with recorded torque were assessed using a patient-clustered Gaussian generalized estimating equation adjusted for sex, age, anatomical sector, implant length and diameter, hard-tissue augmentation, and placement timing. The mean torque was 31.63 ± 5.57 N·cm; the median was 35 N·cm (Q1-Q3, 30-35). Recorded torque was 3.25 N·cm higher for implants placed in women than for those placed in men (95% CI 1.20-5.31; p = 0.002). Relative to maxillary anterior/premolar sites, mandibular molar sites showed higher torque (β = 4.02; p = 0.001), whereas mandibular anterior/premolar sites showed lower torque (β = -5.41; p = 0.013). Implant length had a modest adjusted association (β = 0.93 N·cm/mm; p = 0.022); age, diameter, augmentation, and placement timing did not. The concentration of recorded values at 35 N·cm reflects a clinical target rather than the device limit. These exploratory associations do not establish an intrinsic sex effect or predict osseointegration, and residual confounding due to unmeasured bone characteristics remains possible.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices.Journal of clinical medicine · 2026
- Effect of Progressive Versus Standard Implant Thread Designs on Primary Stability and Insertion Torque: An In Vitro StudyThe International Journal of Oral & Maxillofacial Implants · 2026
- Influence of Implant Macrodesign on Primary Stability Across Different Bone Density Models: A Comparative In Vitro Study.Journal of functional biomaterials · 2026
- In vitro assessment for primary stability of self-drilling orthodontic mini-implants: Pull-out testing at different insertion angles.Bioinformation · 2026
- The Relationship Between Recorded Bone Quality and Primary Implant Stability Measured by Insertion Torque: A Retrospective Study.Cureus · 2026
- Influence of torque-curve dynamics on initial bone remodeling around dental implants: a multicenter prospective study.Journal of periodontal & implant science · 2026