Error Injection in Task-Based Image Quality Pipelines: What Regression Testing Cannot Catch, and Why Neither Internal Identities nor Closed-Form References Suffice Alone.
Shuji Yamamoto
Institute of One, LISIT Co., Ltd., Tokyo 150-0044, Japan.
Key points
From the paper's conclusion- Neither family sufficed alone: the internal identities detected three of the six and the closed-form references five, and peak errors in a reported d' reached 90%.
Abstract
Task-based image quality assessment-the modulation transfer function, the noise power spectrum (NPS), the noise-equivalent quanta and model observers-fails by returning a plausible wrong number rather than an error, and the regression test most implementations carry cannot tell a plausible right answer from a plausible wrong one, because the stored reference was recorded from the defective code. We injected six defects into a validated implementation of that chain through a severity dial that recovers the correct pipeline exactly at zero. A self-consistency regression test detected none of the six. Four internal identities, which need no ground truth, and two closed-form references, which need a phantom whose answer is known, together detected all six, in every case at or before the severity at which the reported detectability index d' became wrong by more than 5%-an error that three of the six never produced. Neither family sufficed alone: the internal identities detected three of the six and the closed-form references five, and peak errors in a reported d' reached 90%. Run unmodified on measured American College of Radiology (ACR) phantom projections across seven reconstruction kernels, the three identities that can be evaluated without ground truth transferred intact-Parseval held to 4×10-16-but their tolerances did not, and the strongest apodisation drove the noise dynamic range to within 0.6% of the threshold beyond which a prewhitening observer should return no value at all rather than a computed one, because 1/NPS has ceased to be numerically meaningful.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- QICOM: Open-Source Software for Reproducible Image Quality Evaluation, Blinded Assessment, and Radiomic ComparisonarXiv · 2026
- It’s not your device, it’s how you use it – Sensitivity and specificity of PSQA devices in the SEAFARER head and neck studyRadiotherapy and Oncology · 2026
- From Augmentation to Synthetic Evidence: Evaluating Generated Chest Radiographs at Dataset IntakearXiv · 2026
- Clinical selectivity and failure modes of automated chest radiograph report evaluation metrics: a cross-dataset analysis of ReXErr-v1 and RadEvalXmedRxiv · 2026
- Epistemic Uncertainty-Aware Defect Detection for Quality Control in Medical Device ManufacturingarXiv · 2026
- MedPhysBench: A Benchmark for Evaluating Language Models on Medical Physics Calculations and Error DetectionarXiv · 2026