One Catheter, Two Chambers: Feasibility and Acute Procedural Outcomes of Dual-Chamber Leadless Pacemaker Implantation with a Single Delivery Catheter.
Marcin Kowalski, Kaiyu Jia, Elizabeth Rimsky, Sean Kowalski, Samer Saouma, Rina Shah, Philippe Akhrass, Valay Parikh
Cardiovascular Institute, Northwell Health, Staten Island University Hospital, New Hyde Park, NY 11040, USA. · Department of Internal Medicine, Staten Island University Hospital, Northwell Health, NY 10305, USA.
Key points
From the paper's conclusion- We recorded acute implant success, catheter malfunction, procedure time, fluoroscopy time, complications through 30 days after implantation, and direct disposable supply cost.
Abstract
Background: Dual-chamber leadless pacing with the Aveir DR system restores atrioventricular synchrony without transvenous leads. The labeled implant workflow calls for a separate delivery catheter for each device. The second catheter adds cost and adds steps to the procedure, and whether it is actually needed has not been examined in a large series. Objective: To describe the acute procedural feasibility, the rate of delivery-catheter malfunction, and the effect on direct disposable supply cost of implanting both devices of a dual-chamber leadless pacemaker system through one delivery catheter. Methods: We reviewed 102 consecutive patients who received a dual-chamber leadless pacemaker at a single center between February 2024 and February 2025, implanted by five operators at that center. Twenty earlier dual-chamber implants performed with two catheters, before the single-catheter workflow was adopted, precede the study period and are not part of this cohort. The right ventricular device was placed first in every case. The same catheter was then reloaded with the atrial device and used for the second implant, and a new catheter was opened only if the first one failed. We recorded acute implant success, catheter malfunction, procedure time, fluoroscopy time, complications through 30 days after implantation, and direct disposable supply cost. Results: Both devices were implanted through the original catheter in 100 of 102 patients (98.0%, 95% CI 93.1 to 99.8%). Two patients (2.0%, 95% CI 0.2 to 6.9%) needed a catheter exchange because of a mechanical failure that arose during ventricular deployment; in both, the ventricular device had been recaptured and redeployed with repeated torque beforehand. No catheter failed during atrial implantation. Mean time from sheath insertion to removal was 69 min, and mean fluoroscopy time was 16.0 ± 6.6 min. Two device dislodgements occurred (2.0%, 95% CI 0.2 to 6.9%), one intraprocedural and one a right atrial device that embolized the day after implant; neither was related to the delivery catheter. There was no perforation, pericardial effusion, or access-site complication, and no other complication occurred in any patient through 30 days after implant. Omitting the second catheter lowered direct disposable supply cost by the cost of the second catheter. Conclusions: With experienced operators, one delivery catheter completed both implants in 98% of cases and lowered disposable supply cost. The study is single arm aand the number of events is small, so these findings support feasibility, with 30-day freedom from complications of 98.0%, and a supply cost difference.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- First short-term European experiences with dual-chamber leadless pacing: Early safety and electrical outcomes compared with transvenous systemsKardiologia Polska · 2026
- Leadless Versus Transvenous Single‐Chamber Ventricular Pacemakers: Real‐World Evidence From Aveir VR Coverage With Evidence Development StudyJournal of the American Heart Association · 2025
- Recommendations on the use of a leadless dual-chamber pacemaker.Cardiology journal · 2026
- Lose the Hardware: The Rise of Leadless Modular Cardiac Pacing.Journal of clinical medicine · 2026
- A decade of leadless pacemaker implantation: technical performance, safety and device longevityEuropean Heart Journal · 2026
- MicraAV Versus Dual-Chamber Transvenous Pacemakers in Patients with Atrioventricular Block Following Transcatheter Aortic Valve Replacement: A 12-Month Single-Center Retrospective Study.Journal of clinical medicine · 2026