Balloon-Expandable Versus Self-Expandable Transcatheter Aortic Valve Implantation Devices: A Contemporary Systematic Review.
Ioannis Apergis, Georgios P Georghiou, Panos Georghiou, Efthalia Emmanouela Chrisofos, Filippos Triposkiadis, Konstantinos Lampropoulos
School of Medicine, European University Cyprus, Nicosia 2404, Cyprus. · Barts and The London School of Medicine and Dentistry, Queen Mary University of London, Turner St., London E1 2AD, UK.
Key points
From the paper's conclusion- Conclusions: Contemporary evidence does not support a universal BEV-versus-SEV hierarchy.
Abstract
Background/Objectives: Transcatheter aortic valve implantation (TAVI) is an established treatment for severe aortic stenosis. Prosthesis selection is increasingly important because balloon-expandable valves (BEVs) and self-expandable valves (SEVs) differ in deployment, hemodynamics, and lifetime-management implications. This systematic review synthesized contemporary evidence comparing BEVs and SEVs, while platform-, generation-, and subgroup-specific studies provided contextual evidence. Methods: A PubMed-only systematic search was performed on 25 January 2026 for studies published between 1 January 2020 and 31 December 2025. Eligible human studies evaluated contemporary balloon or self-expandable systems. Randomized and observational studies, including single-arm cohorts, were included. Risk of bias was assessed using the revised Cochrane Risk of Bias tool (RoB 2) and the Methodological Index for Non-Randomized Studies (MINORS). Because of heterogeneity, findings were synthesized narratively without meta-analysis. Direct randomized and adjusted comparative evidence received greatest weight, whereas single-arm and within-platform studies were not interpreted as establishing comparative superiority. Results: Fifty studies met the eligibility criteria. No consistent mortality advantage was identified for either expansion mechanism. The most reproducible differences concerned hemodynamics, paravalvular regurgitation, and conduction outcomes. Supra-annular SEVs were more consistently associated with lower residual gradients and less prosthesis-patient mismatch, whereas BEVs showed lower paravalvular regurgitation or permanent pacemaker implantation in selected comparative settings. These signals varied by valve generation, implantation technique, anatomy, and study design. Evidence on coronary reaccess, durability, thrombosis, and reintervention was less extensive. Conclusions: Contemporary evidence does not support a universal BEV-versus-SEV hierarchy. Prosthesis selection should be individualized according to anatomy, conduction risk, hemodynamic priorities, and lifetime-management considerations. Comparative conclusions should remain generation-specific and distinguish direct comparative evidence from single-arm or anatomically selected cohorts. The review was not prospectively registered and received no external funding.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Outcomes of Self-Expanding Versus Balloon-Expandable Valves in Severe Aortic Stenosis: A Systematic Review and Meta-Analysis.Cardiology in review · 2026
- Balloon-Expandable Versus Self-Expanding Transcatheter Aortic Valve Implantation in Patients with Bicuspid Aortic Valves: A Systematic Review and Meta-Analysis.Indian heart journal · 2026
- Early Post-transcatheter Aortic Valve Implantation (TAVI) Haemodynamics With SAPIEN Balloon-Expandable and Evolut Self-Expanding Valve Platforms: A Retrospective Single-Centre Cohort Study.Cureus · 2026
- Balloon-expandable valves vs self-expanding valves for transcatheter aortic valve replacement in bicuspid aortic valve stenosis: a meta-analysisEuropean Heart Journal · 2026
- Comparative outcomes of balloon-expandable versus self-expanding transcatheter aortic valves: insights from a large-scale German registry.BMC cardiovascular disorders · 2026
- Long-term outcomes after balloon- or self-expandable transcatheter aortic valve replacement in small aortic annuli.Cardiovascular intervention and therapeutics · 2026